9. SPECIFICATIONS

Principle of measurement:
CH
, CO
;
4
2
Non-dispersion infrared-ray absorption
method
Single light source and single beams
(single beam system)
O
; Fuel cell O
2
H
S ; Constant-potential electrolytic method
2
Measurable gas components and measuring range:
1st range
CH
0 - 20 vol%
4
CO
0 - 20 vol%
2
H
S
0 - 500 ppm
2
O
2
0 - 10 vol%
Built in
fuel cell
each component, and a user can switch
between them.
Measured value indication:
Digital indication in 4 digits
(LCD panel with LED back light)
Analog output signals:
4 to 20mA DC or 0 to 1V DC, isolated
internally from circuit and ground. Output
lines are non-isolated each other.; 12
Digital output: (Option)
1c contact (24V DC/1A, resistive load)
Instrument error, calibration error,
status, solenoid valve drive for auto
calibration, High/Low limit alarm con-
tact output.
* All relay contacts are isolated mutually
and from the internal circuit.
Digital input: (Option)
Voltage contact (supply 12-24VDC (15mA
Remote range change over, auto
calibration remote start, remote hold,
Isolated from the internal circuit with
photocoupler.
Power supply:
Voltage rating
Allowable range
Frequency
Operation conditions:
Ambient temperature ;
5°C to 40°C (CH4, CO, and 0-2000
ppm H
15°C to 40°C (other than the above)
INZ-TN2ZPAF-E
analyzer
2
2nd range
0 - 100 vol%
0 - 100 vol%
0 - 2000 ppm
0 - 5000 ppm
0 - 25 vol%
.
2
; 100V to 240V AC
; 85V to 264V AC
; 50Hz/60Hz
S analyzers)
2
non-condensing
Storage conditions:
Ambient temperature ; -
Dimensions (H × W × D):
Finish color:
Front panel; Cool gray (PANTON 1C-F)
Enclosure:
Steel casing, for indoor use
Material of gas-contacting parts:
Gas inlet/outlet; SUS304
Sample cell; SUS304,chloroprene rubber
Infrared-ray transmitting window; CaF2
ene
Fuel cell O
Gas inlet/outlet: Rc1/4 or NPT1/4 internal thread
N
or air is indispensable)
2
Life time of fuel cell O
2
Life time of constant-potential electrolytic H2S sensor:
1 year (under an average ambient tem-

2. Standard Functions

Output signal holding:
Output signals are held unchanged during
manual and auto calibrations by activation
of holding (turning "ON" its setting).
The values held are those just before
start calibration mode or setting value.
Usage is selectable.
Indication of instantaneous values will
not be held.
Switch ranges:
The switch ranges function is available in
manual, auto, and remote modes. Only
preset switch method is effective.
Manual:
Allows range to switch by key operation.
second range when the measured value
Automatically switched from second range
Remote:
Voltage contact input
signal when remote range switch input
is received.
When the contact input terminals for
each component are input voltage, the
to the second range when the terminals
are open.
* These switch range value are settable between
range values).
71
non-condensing
analyzer cell : ABS resin
2
analyzer: 2 years
-