6.8.2 Span calibration

SPAN
ENT
plied.
Note
For the Ch (components) in which "both" is
set in the calibration range setting (Refer to
Section 6.2.3) - span calibration is completed
for both ranges.
ENT
Note
The indication of span gas changes a few
minutes after you start flowing the span gas.
Keep flowing the gas approximate 10 minutes
until the indication is stabilized, and then carry
out the calibration.
Note: For the Ch (component) for which
"AR" is selected in "6.1.1 Setting
range switch mode," the cursor
automatically moves to the range
selected in "Setting of auto cali-
bration component/range" (6.2.4),
and calibration is carried out
within that range.
To close "Span Calibration"
To close the "Span Calibration" or cancel
this mode midway, press the
A previous screen will return.
SPAN Cal.
key and press
SPAN Cal.
SPAN Cal.
ESC
key.
54
4
sensor.
SPAN
Select Ch No.
with UP / DOWN and ENT
Back with ESC
Ch1
Range 1 0-20.00
CH
Range 2 0-100.0
4
Ch2
Range 1 0-20.00
CO
Range 2 0-100.0
2
Ch3
Range 1 0-500.0
H
S
Range 2 0-2000
2
Range 1 0-10.00
Ch4
O
Range 2 0-25.00
2
(
Select Ch No.
with UP / DOWN and ENT
Back with ESC
Ch1
Range 1 0-20.00
CH
Range 2 0-100.0
4
Ch2
Range 1 0-20.00
CO
Range 2 0-100.0
2
Ch3
Range 1 0-500.0
H
S
Range 2 0-2000
2
Range 1 0-10.00
Ch4
Range 2 0-25.00
O
2
ENT
ENT : Go on calibration
of selected Ch.
ESC : Not calibration
Ch1
Range 1 0-20.00
CH
Range 2 0-100.0
4
Ch2
Range 1 0-20.00
CO
Range 2 0-100.0
2
Ch3
Range 1 0-500.0
H
S
Range 2 0-2000
2
Ch4
Range 1 0-10.00
Range 2 0-25.00
O
2
ENT
To Measurement screen after
executing Manual Span Calibration
vol%
0.0
vol%
vol%
0.0
vol%
ppm
0.00
ppm
0.0
vol%
vol%
ENT
)
vol%
0.0
vol%
vol%
0.0
vol%
ppm
0.00
ppm
vol%
0.0
vol%
vol%
0.0
vol%
vol%
0.9
vol%
ppm
0.34
ppm
1.1
vol%
vol%
INZ-TN2ZPAF-E

Advertisement

Table of Contents
loading

Table of Contents