6.8 Calibration

6.8.1 Zero calibration

ZERO
Note
For the Ch (components) in which "both" is
set in the calibration range setting (see Section
6.2.3) - zero calibration is carried out on both
ranges.
Note1: For the Ch (component) for
which "AR" is selected in "Sec-
tion 6.1.1 Setting range switch
mode," the cursor automatically
moves to the next range selected
in "Setting of auto calibration
component/range" (Section
6.2.4), and calibration is carried
out within that range.
Note2: Zero point for O
is calibrated to 21% at zero
calibration, because this gas
analyzer uses air as zero gas.
To close "Zero Calibration"
To close the "Zero Calibration" or cancel this mode midway, press the
A previous screen will return.
INZ-TN2ZPAF-E
ENT
ENT
measurement
2
53
ZERO
ZERO Cal.
Select Ch No.
with UP / DOWN and ENT
Back with ESC
Ch1
Range 1 0-20.00
CH
Range 2 0-100.0
4
Ch2
Range 1 0-20.00
CO
Range 2 0-100.0
2
Ch3
Range 1 0-500.0
H
S
Range 2 0-2000
2
Range 1 0-10.00
Ch4
Range 2 0-25.00
O
2
(
ZERO Cal.
Select Ch No.
with UP / DOWN and ENT
Back with ESC
Ch1
Range 1 0-20.00
CH
Range 2 0-100.0
4
Ch2
Range 1 0-20.00
CO
Range 2 0-100.0
2
Ch3
Range 1 0-500.0
H
S
Range 2 0-2000
2
Range 1 0-10.00
Ch4
Range 2 0-25.00
O
2
ENT
ZERO Cal.
ENT : Go on calibration
of selected Ch.
ESC : Not calibration
Ch1
Range 1 0-20.00
CH
Range 2 0-100.0
4
Ch2
Range 1 0-20.00
CO
Range 2 0-100.0
2
Ch3
Range 1 0-500.0
H
S
Range 2 0-2000
2
Range 1 0-10.00
Ch4
Range 2 0-25.00
O
2
ENT
To Measurement screen after
executing Manual Zero Calibration
vol%
0.0
vol%
vol%
0.0
vol%
ppm
0.00
ppm
vol%
0.0
vol%
)
vol%
0.0
vol%
vol%
0.0
vol%
ppm
0.00
ppm
vol%
0.0
vol%
vol%
0.0
vol%
vol%
0.9
vol%
ppm
0.34
ppm
vol%
1.1
vol%
ESC
key.