Fuji Electric ZPAF Instruction Manual page 40

Biomass gas analyzar
Table of Contents

Advertisement

To close the setting
ESC
Press the
key to exit automatic calibration component/range setting, and the previous
screen appears.
Operation by setting
Auto calibration is performed under the following rules.
1. Zero calibration is performed at the same time, for the Ch (component) in which "enable"
is selected at the time of auto calibration and auto zero calibration.
2. Span calibration is performed in the order from smallest Ch No., for the Ch (component)
for which "enable" is selected at the time of auto calibration.
Note
ZERO calibration on auto calibration and auto zero calibration of the component for
which "enable" is selected are performed in batch irrespective of the description in
"6.2.2 Setting of manual zero calibration."
INZ-TN2ZPAF-E
29