Remote Start - Fuji Electric ZPAF Instruction Manual

Biomass gas analyzar
Table of Contents

Advertisement

cases.
Example
Start Time
Cycle
Flow Time
ON/OFF
In case where auto calibration is carried out at the above setting.
Sunday
12:00
Cycle
ChI
Zero
Span
calibration
calibration
350sec
350sec
Flow time
(An example of "Ch1: through Ch5: enable", as given in Section 6.2.4 "Auto Calibration
Components/range")
Setting range
Cycle
: 1 to 99 hours or 1 to 40 days (initial value 7 days)
Flow time
: 60 to 900 sec
Caution
When an auto calibration starts, the measurement screen appears automatically.
Any operation other than "Stop Auto Calibration" (see Section 6.4.2) is
not permitted during auto calibration. "Stop Auto Calibration" cannot be performed
with the key lock to ON. To cancel auto calibration forcedly, set the key lock to OFF
and then execute "Stop Auto Calibration".
Turn on the power again after it is turned off (including the case of power failure) at
the time set as the next start time in auto calibration, and then repeat it in the set cycle.

Remote start

Remote start input
INZ-TN2ZPAF-E
12:00
SUN
day
1
350 sec
Zero
350 sec
Ch1 Span
350 sec
Ch2 Span
350 sec
Ch3 Span
300 sec
Ch4 Span
300 sec
Ch5 Span
300 sec
EX. time
ON
Monday
12:00
Ch2
Ch3
Ch4
Span
Span
Span
calibration
calibration
calibration
350sec
350sec
300sec
(initial value 300 sec)
Tuesday
12:00
Ch5
Replace-
Span
ment time
calibration
300sec
300sec
With input (hold at least 1.5 sec.)
Without input
35
: Auto calibration