6.2.3 Setting of calibration range

To close "Setting of Calibration Range"
To close "Setting of Calibration Range" or to cancel this mode midway, press the
A previous screen will return.
Example
Ch1
Range 1: 0 to 20 vol%
CH
Range 2: 0 to 100 vol%
4
Ch2
Range 1: 0 to 20 vol%
CO
Range 2: 0 to 100 vol%
2
Ch1: Range 1 and Range 2 are calibrated together.
Ch2: Only currently displayed range is calibrated.
Note
Zero point for O
measure-
2
ment is calibrated to 21% at
zero calibration, because this
gas analyzer uses air as zero
gas.
INZ-TN2ZPAF-E
key.
ENT
When setting CH
Two cursors will appear in both ranges (Ch1 and Ch4).
Cell. Settings
Cell. Range
Ch1
Range1
CH
Range2
4
Ch2
Range1
ENT
CO
Range2
2
Ch3
Range1
H
S
Range2
2
Ch4
Range1
Range2
O
2
End of Calibration Range Setting
Manual Calibration screen
and O
4
2
ZERO Call.
ENT : Go on Calibration
of selected Ch
ESC : Not calibration
Ch1
Range1
CH
Range2
4
Ch2
Range1
CO
Range2
2
Ch3
Range1
H
S
Range2
2
Ch4
Range1
O
Range2
2
27
Set calibration range
current or both range
0–20.00
vol%
both
0–100.0
vol%
0–20.00
vol%
current
0–100.0
vol%
0–500.0
ppm
current
0–2000
ppm
0–10.00
vol%
both
0–25.00
vol%
ENT
(
)
ESC
key.
both
current
to "both"
-0.6
0–20.00
vol%
0–100.0
vol%
0.4
0–20.00
vol%
0–100.0
vol%
0.00
0–500.0
ppm
0–2000
ppm
-0.1
0–10.00
vol%
0–25.00
vol%