Setting Of Calibration Range - Fuji Electric ZPB Insruction Manual

Ndir type infrared gas analyzer
Hide thumbs Also See for ZPB:
Table of Contents

Advertisement

6.2.3 Setting of calibration range

ENT
To close "Setting of Calibration Range"
To close "Setting of Calibration Range" or to cancel this mode midway, press the
A previous screen will return.
Example
Ch1
Range 1: 0 to 50 ppm
NOx
Range 2: 0 to 500 ppm
Ch2
Range 1: 0 to 50 ppm
SO
Range 2: 0 to 500 ppm
2
Ch1: Range 1 and Range 2 are calibrated together.
Ch2: Only currently displayed range is calibrated.
Note
To perform calibration for
"both," set the same calibra-
tion gas concentration for
both ranges.
INZ-TN2ZPB-E
ENT
key.
When setting NOx and CO to "both"
Two cursors will appear in both ranges (Ch1 and Ch4).
Cell. Settings
Cell. Range
Ch1
Range1
-
NOx
Range2
Range1
Ch2
SO
Range2
2
Ch3
Range1
CO
Range2
2
Ch4
Range1
CO
Range2
Ch5
Range1
-
Range2
O
2
-
End of Calibration Range Setting
Manual Calibration screen
ZERO Call.
ENT : Go on Calibration
of selected Ch
ESC : Not calibration
Ch1
Range1
0–50.00
NOx
Range2
0–500.0
Range1
0–50.00
Ch2
SO
Range2
0–500.0
2
Ch3
Range1
0–10.00
CO
Range2
0–20.00
2
Ch4
Range1
0–50.00
CO
Range2
0–500.0
Ch5
Range1
0–10.00
Range2
0–25.00
O
2
31
Set calibration range
current or both range
0–50.00
ppm
both
0–500.0
ppm
0–50.00
ppm
current
0–500.0
ppm
0–10.00
vol%
current
0–20.00
vol%
0–50.00
ppm
both
0–500.0
ppm
0–10.00
vol%
current
0–25.00
vol%
ENT
(
)
ESC
key.
both
current
-0.6
ppm
ppm
0.4
ppm
ppm
0.00
vol%
vol%
-0.1
ppm
ppm
vol%
21.00
vol%