6.2.3 Setting of calibration range

ENT
To close "Setting of Calibration Range"
To close "Setting of Calibration Range" or to cancel this mode midway, press the
A previous screen will return.
Example
Ch1
Range 1: 0 to 10.00 vol%
CO
Range 2: 0 to 20.00 vol%
2
Ch2
Range 1: 0 to 500.0pm
CO
Range 2: 0 to 2000 ppm
Ch1: Range 1 and Range 2 are calibrated together.
Ch2: Only currently displayed range is calibrated.
Note
To perform calibration for
"both," set the same calibra-
tion gas concentration for
both ranges.
ENT
key.
When setting Ch1 (CO
Two cursors will appear in both ranges (Ch1).
Cal. Settings
Cal. Range
Ch1
Range1
-
CO
Range2
2
Ch2
Range1
CO
Range2
-
-
End of Calibration Range Setting
Manual Calibration screen
) to "both"
2
ZERO Cal.
Ch1
Range1
CO
Range2
2
Ch2
Range1
CO
Range2
28
Set calibration range
current or both range
0–10.00
vol%
both
0–20.00
vol%
0–500.0
ppm
current
0–2000
ppm
ENT
(
)
ESC
key.
both
current
ENT : Go on calibration
of selected Ch
ESC : Not calibration
0–10.00
0.00
vol%
0–20.00
0.00
vol%
0–500.0
0.0
ppm
0–2000
ppm
INZ-TN2ZPAH-E