6.2.3 Setting of calibration range

To close "Setting of Calibration Range"
To close "Setting of Calibration Range" or to cancel this mode midway, press the
A previous screen will return.
Example
Note
To perform calibration for
"both," set the same calibra-
tion gas concentration for
both ranges.
key.
ENT
When setting NOx to "both"
Two cursors will appear in both ranges (Ch1).
Cal. Settings
Cal. Range
Ch1
Range1
NOx
Range2
Ch2
Range1
ENT
O
Range2
2
End of Calibration Range Setting
Manual Calibration screen
ZERO Cal.
ENT : Go on calibration
of selected Ch
ESC : Not calibration
Ch1
Range1
NOx
Range2
Ch2
Range1
O
Range2
2
42
Set calibration range
current or both range
0–10.00
ppm
both
0–100.0
ppm
0–10.00
vol%
current
0–25.00
vol%
(
)
ENT
ESC
key.
-0.6
0–10.00
ppm
0–100.0
ppm
0–10.00
vol%
21.00
0–25.00
vol%