6.2.3 Setting of calibration range

ENT
To close "Setting of Calibration Range"
To close "Setting of Calibration Range" or to cancel this mode midway, press the
A previous screen will return.
Example
Ch1
Range 1: 0 to 500 ppm
NOx
Range 2: 0 to 5000 ppm
Ch2
Range 1: 0 to 500 ppm
Range 2: 0 to 2500 ppm
SO
2
Ch1: Range 1 and Range 2 are calibrated together.
Ch2: Only currently displayed range is calibrated.
Note
To perform calibration for
"both," set the same calibra-
tion gas concentration for
both ranges.
ENT
key.
When setting NOx and O
Two cursors will appear in both ranges (Ch1 and Ch4).
Cal. Settings
Cal. Range
Ch1
Range1
-
NOx
Range2
Ch2
Range1
SO
Range2
2
Ch3
Range1
CO
Range2
Ch4
Range1
O
Range2
2
-
-
End of Calibration Range Setting
Manual Calibration screen
2
ZERO Cal.
Ch1
Range1
CO
Range2
2
Ch2
Range1
SO
Range2
2
Ch3
Range1
CO
Range2
Ch4
Range1
O
Range2
2
32
Set calibration range
current or both range
0–500.0
ppm
both
0–5000
ppm
0–500.0
ppm
current
0–2500
ppm
0–200.0
ppm
current
0–2000
ppm
0–10.00
vol%
both
0–25.00
vol%
ENT
(
)
ESC
key.
both
current
to "both"
ENT : Go on calibration
of selected Ch
ESC : Not calibration
0–500.0
-0.6
ppm
0–5000
ppm
0–500.0
0.4
ppm
0–2500
ppm
0–200.0
0.00
ppm
0–2000
ppm
-0.1
0–10.00
vol%
0–25.00
vol%
INZ-TN2ZPAJ-E