Rest profile (manual pull)
Press the Start Pull button to mark the start of the profile (note: this is
an immediate marker). The metronome will help you to pull the catheter
with a constant speed. One beep corresponds to one centimeter. An
increase in urethral pressure (P
enters the urethra. When P
marker will be inserted automatically (backward in time). The position
of the marker will be determined by the software and is based on P
at the start. When P
marker will be inserted automatically by the software. Also, the
Maximum Urethral Closure Pressure marker (MUCP) will be inserted
automatically on the highest pressure.
Press the End pull button to mark the end of the profile. Position both
sensors (P
Stress profile (manual pull)
Press the Start Pull button to mark the start of the profile (note: this is
an immediate marker). The metronome will help you to pull the catheter
with a constant speed. One beep corresponds to one centimeter. Ask
the patient to cough every two seconds. The peaks caused by
coughing are used to calculate transmissions. An increase in urethral
pressure (P
When P
ura
automatically (backwards in time). The position of the marker will be
determined by the software and is based on P
decreases to a certain value, a Profile end marker will be inserted
automatically by the software. Also, the Maximum Urethra Closure
Pressure marker (MUCP) will be inserted automatically on the highest
pressure.
Press the End pull button to mark the end of the profile. Position both
sensors (P
96
ura
decreases to a certain value, a Profile end
ura
and P
) in the bladder.
ves
ura
) indicates that the urethral sensor enters the urethra.
ura
increases > 10 cmH
and P
) in the bladder.
ves
ura
Nexam Pro User Manual
) indicates that the urethral sensor
ura
increases > 10 cmH
O, a Profile start marker will be inserted
2
ura
O, a Profile start
2
at the start. When P
Chapter 7
ura
ura