o Water catheters (see § 3.6).
Connect the pump tube to the filling lumen of the catheter.
Prepare the EMG cables and apply the EMG electrodes (see § 2.5.10).
Press the Start investigation button.
Press the Pump ON button to start the pump.
Press the First sensation button to mark the first sensation. Use this
button for all sensation markers.
When the patient is unable to tolerate further filling or if the bladder is filled
sufficiently, press the Pump button to stop.
Press the Stop investigation button. Enter the residual urine if prompted.
Remote control buttons
Buttons
78
Description
Press Shift and Zero all simultaneously to zero the
detrusor pressure (P
necessary, during the measurement.
Press Valsalva begin/end to start/end the Valsalva
manoeuvre.
Press Shift and Leakpoint to start the Cough Leak Point
Pressure procedure (CLPP). Ask the patient to cough; the
software will insert cough markers around the peaks.
Press Leakpoint to mark a leakpoint.
Nexam Pro User Manual
) during the pre-test and, if
det