MEMORY FULL **** PRETEST DATA IS NOT
SAVED NOW!
Microsoft Bluetooth stack or Widcomm
Bluetooth stack not installed. Use Windows XP
with at least Service Pack 2 to use the Microsoft
Bluetooth stack or re-install the software to
install the Widcomm Bluetooth stack.
LABORIE perfusion pressure too high - The
pressure release valve is probably polluted.
Release the tube from the water container and
try to restart the pump. If the error remains, the
system should be sent to LABORIE for
maintenance. Restart or Stop and save
investigation?
LABORIE perfusion pressure too high. The
pressure release valve is probably polluted.
Release the tube from the water container and
try to restart the pump. If the error remains, the
system should be sent to LABORIE for
maintenance. You can restart the perfusion
pump with the fill perfusion button.
LABORIE perfusion pressure too low. Check
Leakage. Please make sure all cables and
connections on the water container are airtight.
No esophageal channels defined!
No Input signal present!
No Luna connected.
NPU not connected!
Oximeter is not active or within range or the
COM port setting is wrong. Retry connecting?
Parameter file could not be saved!
Patient demographics could not be written!
Perfusion system with balloon filling not
connected!
Protocol could not be found
Protocol file is corrupt. Open Protocol editor to
fix this problem.
Protocol file is corrupt. Protocol
[ProtocolName], [TestName] Open Protocol
editor to fix this problem.
Puller error
Solar Blue not connected!
Test not present!
Software Messages
Cause
Memory full, Pretest took longer
than expected
Communication problems
Perfusion problem
Perfusion problem
Perfusion problem
No Esophageal channel defined
No video input signal available
Module expected but not
connected
Module expected but not
connected
Module not active
Error during saving of parameter
file
Error during saving of patient
information
Module expected but not
connected
Protocol file is corrupt
Protocol file is corrupt
Protocol file is corrupt
Puller not running at correct speed
or not at all
Module expected but not
connected
Investigation file is not available
Nexam Pro User Manual
Posible actions to
resolve
Stop the investigation
and start a new
investigation if
applicable
Contact your LABORIE
representative
Contact your LABORIE
representative
Contact your LABORIE
representative
Contact your LABORIE
representative
Define esophageal
channels
Contact your LABORIE
representative
Contact your LABORIE
representative
Contact your LABORIE
representative
Retry connecting
Contact your LABORIE
representative
Contact your LABORIE
representative
Contact your LABORIE
representative
Create new protocol
Open protocol editor to
fix this problem
Open protocol editor to
fix this problem
Check puller or contact
your LABORIE
representative
Contact your LABORIE
representative
Contact your LABORIE
representative
265