LABORIE Nexam Pro WPU-L4 User Manual page 101

Table of Contents

Advertisement

Resting bladder pressure
The resting bladder pressure is the urethra pressure at the position of the
profile begin marker. It is only calculated when the profile begin marker is
present.
At the moment the profile starts (just before the urethral pressure increases
because the catheter reaches the beginning of the urethra), the transducer
measuring the urethral pressure Pura is still in the bladder; hence, this is the
resting bladder pressure.
Calculated: Resting bladder pressure = [F].
Maximum Urethral Closure Pressure (MUCP)
Calculation method
The Maximum Urethral Closure Pressure (MUCP) can be calculated in two
ways:
• MUCP = Max P
• MUCP = Max (P
In the analysis program, you can define how to calculate the Maximum
Closure Pressure. Choose Settings > UPP settings from the menu.
Figure 7.4 MUCP calculation method
MUCP = Max P
ura
MUCP is the difference between the maximum urethral pressure and the
resting bladder pressure.
Max P
is the maximum urethral pressure (P
ura
marker and the Profile end marker. P
This is the pressure at the position of the Profile start marker on the urethral
channel (P
). At the moment the profile starts (just before the urethral
ura
Urethral Pressure Profile (UPP)
– P
(ura)
ura
ves
– P
) default
ura
ves
– P
(ura)
ves
Nexam Pro User Manual
) between the Profile start
ura
(ura) is the resting bladder pressure.
ves
101