Filling phase (continue)
Press the Pump button to start the pump.
You can mark the sensations. At different filling volumes you may stop
the pump to perform rest and stress profiles.
When the patient is unable to tolerate further filling or if the bladder is
filled sufficiently, press the Pump button to stop.
Press the End filling phase button to place the Storage phase marker.
You can do different tests now to provoke leakage. Press the Leak
point button to mark leakage.
Press the Voiding phase button to mark end filling phase. The
investigation is now temporarily halted. Reposition the patient if
necessary.
Voiding phase
Press the Start voiding phase button.
Press the Permission to void button to mark this command and then
leave the patient alone. The LED on the remote control (RC-1600) will
provide you with feedback when the uroflow has been finished. Also,
the buzzer of the remote control will beep after the flow has stopped.
(You can find more information in § 12.3.4).
If wanted, you can activate the Tap water sound during miction. Click
the button again to deactivate.
Press the Stop investigation button. Enter the residual urine if
prompted.
Gynecology Study
Nexam Pro User Manual
109