Press the Balloon expulsion End button to mark the end of the test.
The software asks whether the balloon expulsion was successful.
Select Failed, Partial, or Success and then click OK.
Figure 9.20 Balloon expulsion test
Filling with water (pump module)
Press the Next test button until the text Balloon expulsion is
displayed on the left button. Connect the filling lumen of the catheter to
the filling tube that is placed on the pump of the Nexam Pro. The filling
speed is displayed on the Pump button in the upper-left corner of the
screen (e.g., 10 ml).
Press the Pump button to start the pump. The balloon in the rectum
will be slowly filled.
When the desired volume is reached, stop the pump by pressing the
Pump button.
Press the Balloon expulsion button to mark the start of the test. A
balloon marker is automatically inserted. Ask the patient to expel the
balloon.
Press the Balloon expulsion End button to mark the end of the test.
The software asks if the balloon expulsion was successful. Select
Failed, Partial, or Success and then click OK.
Filling with air (Perfusion Pump Plus)
Anorectal Manometry
Nexam Pro User Manual
141