LABORIE Nexam Pro WPU-L4 User Manual page 267

Table of Contents

Advertisement

Message
the pump. If the error remains, the system
should be sent to LABORIE for maintenance.
Warning: compliance marker in miction phase
Warning: No more Profile markers!
Warning: No more Profile markers!
Wireless patient module connection lost.
WPM not connected! Check the Range of the
Wireless patient module, Check the Bluetooth
dongle or replace the batteries
WPM not connected! Check the Range of the
Wireless patient module, Check the Bluetooth
dongle or replace the batteries
WPU not connected! Check the Range of the
Wireless patient unit, Check the Bluetooth
dongle or replace the batteries
WPU not connected! Check the Range of the
Wireless patient unit, Check the Bluetooth
dongle or replace the batteries
Writing of channel [Channel name] to catheter
file failed!
Software messages database
Message
Can not execute [application]
Chipcard error
Could not open combination report because
of a problem with following investigation:
[Report file] Generate report file by re-
opening the investigation and create a new
combination report.
Creating a backup also failed. All changes
made have gone lost. Contact your
administrator.
Creating safety backup failed!
Database could not be reindexed
Software Messages
Cause
Compliance marker in miction
phase
No more profile markers available
No more profile markers available
Communication problems
Module expected but not
connected
Module expected but not
connected
Module expected but not
connected
Module expected but not
connected
Writing to catheter file failed
Cause
Application cannot be executed Contact your LABORIE
Chipcard could not be read
N/A
Failed to create a backup
Safety backup could not be
copied
N/A
Nexam Pro User Manual
Posible actions to
resolve
Remove marker
Remove profile marker
Remove profile marker
Contact your LABORIE
representative
Contact your LABORIE
representative
Contact your LABORIE
representative
Contact your LABORIE
representative
Contact your LABORIE
representative
Retry calibration
Posible actions to
resolve
representative
Contact your LABORIE
representative
Contact your LABORIE
representative
Contact your LABORIE
representative
Software will
continue to work as
normal without the
safety backup,
contact your LABORIE
representative
Contact your LABORIE
representative
267

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Nexam pro wpu-esNexam pro cim

Table of Contents