Safety Information
60601 system requirements should be applied prior to
use.
Do not use High-Frequency (HF) surgical equipment on
the patient or in the vicinity of the Nexam Pro system as
the patient may experience burns. Device is not
intended for use with HF surgical equipment.
Do not place the safety module/Multiple
Socket-Outlet (MSO) on the floor.
Do not connect an additional Multiple Socket-
Outlet (MSO) or extension cord to the Nexam
Pro system.
Do not connect non-ME equipment that is
supplied as part of this system intended to be
plugged into the Separating transformer
module into a wall outlet. The Separating
transformer is a safety component designed to
ensure Touch/Enclosure Leakage currents are
below the allowable limits. Operator may
experience electrical shock if non-ME
equipment is connected directly to the wall
outlet.
Only connect equipment that has been
supplied as part of the Nexam Pro system to
the Multiple Socket Outlet provided on the
Separating transformer.
The digital MMS-bus cables connected to the Nexam
Pro modules all have a connector with locking
mechanism. Do not attempt disconnection by pulling the
cable. Loosen the lock by pressing the lever on the
connector.
Nexam Pro User Manual
243