Page 2
Trademarks NXT PRO is a trademark of LABORIE Medical Technologies Canada ULC. Roam NXT, PIM NXT, Urocap NXT, and Urocap NXT are trademarks of LABORIE Medical Technologies Canada ULC. T-DOC is a trademark of T-DOC Company, LLC. Unisensor is a trademark of Unisensor AG. Windows is a registered trademark of Microsoft Corp.
Contents Introduction ...................... 12 Intended Use ....................12 Indications of Use ..................12 Equipment Intended Use ................13 Contraindications ..................13 About this Manual ..................14 Manual Symbols ..................14 Software and Equipment ................14 Software Install ..................14 Standard Equipment and Optional Accessories ........... 15 Warnings and Precautions ................
Page 4
3.3.2.2 Preview 2 Phase ................. 31 3.3.2.3 Filling Phase ..................32 3.3.2.4 Voiding Phase ..................33 3.3.2.5 Urology Urodynamics Report ..............33 Cystometry Study ..................34 Starting the Cystometry Study ..............34 Performing the Cystometry Study ............34 3.4.2.1 Preview phase ..................34 3.4.2.2 Cystometry phase ................
Page 9
List of Figures Figure 1: NXT Pro System Cart .................. 12 Figure 2: Lockable Wheels ..................18 Figure 3: Roam NXT and Urocap NXT inserted in NXT Docking Station ......19 Figure 4: Connecting the NXT Pro Cart to Supply Mains ..........19 Figure 5: Urocap NXT Setup ..................
Page 10
Figure 47: Manual Consumable Traceability ..............59 Figure 48: Change Channel Color ................59 Figure 49: Configuration Settings – Editing the Report ..........61 Figure 50: Comments Window ................... 61 Figure 51: Comments and Header Title Drop Down Menus ..........62 Figure 52: Normal Values ..................
Page 11
List of Tables Table 1: NXT System Configuration – Standard Equipment ..........15 Table 2: Synergy Main Window Section Overview ............42 Table 3: Title Bar Icon Overview ................42 Table 4: Settings Options Overview ................43 Table 5: Admin Options Overview ................43 Table 6: Synergy Navigation Icons ................
Introduction Introduction The LABORIE NXT system is a fully wireless capable system providing mobility and ease-of- use in the field of Urodynamics. It allows studies to be performed with mountable modules and is controlled by the LABORIE Synergy software. The wireless and battery-powered features of each device allow for improved system setup and use.
Introduction 1.3 Equipment Intended Use Refer to the descriptions below for a breakdown of intended use per module. Standard Equipment • NXT Docking Station with Pump: Wired device which uses a peristaltic infusion pump to fill a patient’s bladder with saline and/or contrast from an IV bag. •...
Microsoft Windows ® . The UDS Urodynamics system and software provided by LABORIE are tested as a complete package and will not work correctly after updating the operating system. Contact your LABORIE support representative for more information. Changes to the Windows Group Policies may prevent the NXT system from •...
Standard Equipment and Optional Accessories Prior to use, verify that all ordered equipment and accessories have been received. Contact LABORIE if there are any discrepancies between ordered and received equipment. Inspect the equipment for any visible signs of damage or mishandling. Notify the carrier immediately if damage is found.
Page 16
Introduction LABORIE is not responsible for loss of patient files or test data. LABORIE recommends that the user backup patient data on a regular basis. To prevent any risk of damage to the cart: Ensure the monitor is free of any obstacles.
Page 17
EMC interference. Minimize interference from other Bluetooth devices by setting up all modules of the system in proximity to each other. Only connect devices specified and approved by LABORIE to the NXT System multiple socket outlets. Connecting any device not approved by LABORIE to the Medical Electrical (ME) system alters the functional integrity of the product creating a safety risk.
(where applicable) DVI to BNC adaptor. CAUTION: Consult LABORIE Service before connecting any video device to the NXT System. Anyone connecting supplementary equipment to ME equipment configures a medical system and is therefore responsible for ensuring that the system complies with the requirements for medical electrical systems, encompassed in the IEC/EN 60601 series.
Quick Start NXT Docking Station Urocap NXT Charging Hub Figure 3: Roam NXT and Urocap NXT inserted in NXT Docking Station The NXT Pro Cart’s power cable plugs in at base of the NXT system. Connect the NXT system power cord to the NXT Pro Cart as shown in the figure below, then plug the NXT Cart’s power cable to an external power source.
Quick Start Turn on the AIO PC by pressing the power button . Double click the Synergy Icon to launch the Synergy Login Page. Login to Synergy by inputting the account User Name and Password, assigned during system install, into the Synergy Login Page. The Synergy Main Window will load.
IMPORTANT: Ensure all preparation steps are completed before starting a study in Synergy. 2.4 Preparing the Roam NXT Ensure the Roam NXT is charged. LABORIE recommends charging the device • overnight to ensure optimal battery power for a full day of procedures.
Figure 6: Consumable Traceability – Preview Window Infusion Pump Tubing Setup LABORIE Infusion Pump Tubing is intended for single use only. Do not re-use, reprocess, or resterilize pump tubing. Use of the Infusion Pump Tubing outside of its intended use may affect pump accuracy.
Quick Start To setup the pump tubing, follow the steps provided below: Scan the SmartSense tag, located on the Tubing packaging, against the reader located on the NXT Docking Station by holding proximity for 3 seconds. Refer to NXT Docking Station section on page for a visual of the SmartSense tag reader.
Quick Start For female patients: Insert vesical (Pves) catheter 8-10 cm for single sensor, 12-14 cm for dual sensor. Insert abdominal (Pabd) catheter into the rectum 10-15 cm, past any stool that may be present, along the anterior wall of the rectum. Alternatively, insert vaginally in the posterior fornix, just behind the cervix, at the level of the cul-de-sac of Douglas.
Quick Start IMPORTANT: TDOC Air-Charged Catheters are intended for single-use. Synergy will not recognize a catheter that has previously been registered in the software after the allotted two hours to complete Urodynamic testing. Once preferred catheter placement is confirmed, connect the Infusion Pump Tubing to the filling lumen on the vesical catheter or filling catheter.
How to Perform a Study How to Perform a Study To perform a typical study, follow the instructions provided. Starting a Study Running the Study FreeFlow Study Urology Urodynamic Study iii) Cystometry Study Tests Finishing the Studies Printing the Study Result (Optional) 3.1 Starting a Study Ensure that the computer and printer are turned on.
How to Perform a Study b) In the Device Manager window, identify the Bluetooth Dongle NXT, Urocap NXT, Roam NXT, and Pump NXT and verify each device is properly connected as indicated by a green checkmark. c) To connect the Remote Control NXT, ensure the remote control is active then click the Search for devices button.
A FreeFlow study provides a measurement of the rate at which urine flows out of the body. It can be performed using the Urocap NXT and Synergy software provided as part of the LABORIE NXT system. NOTE: Before beginning the FreeFlow Study ensure that you have completed the tasks...
How to Perform a Study Click the Zero Volume button to prepare the Urocap NXT and then click the Start Flow Phase button to start the study and move to the Flow phase. NOTE: Pause the study by clicking the Pause button and resume the study by clicking the resume button.
How to Perform a Study Figure 16: Post Voided Residual Window Click OK on the Post Void Residual window to save data and advance to the Report. Pressing Cancel skips the PVR window. NOTE: Post Voided Residual may also be entered and edited in the Report. 3.2.2.3 FreeFlow Report A preview of the FreeFlow Report will be generated using the default (UDS)
How to Perform a Study 3.3 Urology Urodynamic Study The Urology Urodynamic Study is a urologic diagnostic test that provides aid in the diagnosis of urine storage and voiding disfunction. The Urology Urodynamic Study includes a FreeFlow test and a Pressure Flow test. A Pressure Flow test involves filling the bladder and determining the Detrusor pressure Pdet = Pves –...
How to Perform a Study 3.3.2.3 Filling Phase Click the Start Filling Phase button . The Filling phase window will display. NOTE: Pause the study by clicking the Pause button and resume the study by clicking the resume button. If necessary, the Pabd channel value can be set to equal the Pves channel value by pressing the Equalize Pdet button.
How to Perform a Study Caution: Infusion can be stopped at ALL TIMES by pressing the grey pump stop button on the NXT Docking Station, by opening the pump door, or by disconnecting the filling tube from the catheter or the water bag. 3.3.2.4 Voiding Phase The Voiding phase will begin (Figure 19).
How to Perform a Study 3.4 Cystometry Study A Cystometry study involves filling the bladder and determining the Detrusor pressure Pdet = Pves – Pabd (the difference between the pressure inside the bladder (Pves) and the pressure inside the abdomen (Pabd)). The Cystometry study provides one phase to complete filling and voiding.
How to Perform a Study selected in error, delete the event and continue “filling phase” activities by pressing the Pump On button. The Cystometry phase will stop 60 seconds after voiding ends. Alternatively click the Finish Procedure button. A Post Void Residual window will pop-up. Select the preferred method of capturing Post Voided Residual (PVR), meaning the amount of fluid remaining in the bladder at the completion of the voiding phase.
Print icon. 3.8 Troubleshooting For FAQs and troubleshooting tips, refer to the Troubleshooting Guide section on • page 88. If problems cannot be resolved, contact the LABORIE Service team at 1- 800-333-1039 or email service@laborie.com. NXT Owner’s Manual NXT-UM01, MAN2010...
To login to Synergy, follow the steps below. Double-click the Synergy shortcut icon on the desktop or launch the Synergy application from the LABORIE Dashboard. The Synergy Login Page will be displayed (Figure 21). Upon first login ensure to accept the End User License Agreement.
41. IMPORTANT: Do not leave the computer unattended once login is completed. LABORIE recommends that the user manually lock the computer before leaving the computer workstation. Creating User Accounts Synergy facilitates four basic account structures: Local Admin, Local User, Domain Admin, and Domain User.
Synergy Software Figure 23: User Setup Window To reset the password of an active account, select the existing account from the user list. Under the Update User Info tab, enter the new password and press OK (Figure 24). Figure 24: Password Reset NOTE: Local and Domain Users may reset their own passwords ONLY.
Synergy Software 4.1.2.2 Managing User Role All Synergy account types have pre-set permissions that control which function users may access based on user roles. Local and Domain Admins may manage user role permissions with the Manage User Role window. Follow the instructions provided below for managing user roles.
Synergy Software Figure 27: Adding New User Role To reassign the account role, click the Admin icon located on the Synergy title bar and select User Setup > Update User Info Tab. Select the newly created role; press Save. 4.2 Sections and Icons in the Main Window This section overviews the various features accessible from the Main Window.
Synergy Software Overview of Sections A brief overview of the sections in the Main Window is provided in Table 2. Refer to Figure 28 for a visual guide showcasing the location of all sections described. Section Name Section Description Patients Lists all the patients.
Synergy Software Database Contains Archive Patients, Import Patients, and Deleted Management Patients options. Contains Editor, Import Custom Workflow, and Export Custom Workflow Workflow. Editor displays the Workflow Configuration page; create new workflows and set hardware channels. Table 4: Settings Options Overview 4.2.2.2 Admin The options under the Admin icon are described in Table 5 below.
Synergy Software 4.3.1.1 Adding a New Patient Click the New Patient button in the Patients section. Enter the patient details in the Patient Details section under the Info tab (Figure 30). Click the Save button under the Info tab to save the patient details. If required, enter the relevant past medical history under the History tab.
Synergy Software 4.3.1.4 Exporting Patient Data Patient data can be exported from the system in different forms (e.g. PDF, HTML, RTF, DOCX, XLS, XLSX, Image) to a computer. Follow the instruction provided below to export patient data. Select the patient from the Patents section. Open the report from Patient Details.
Patient Display Settings. By default, the EMR field will be off. Select Table in the table options column and click Save (Figure 33). Figure 33 Patient Settings – EMR NOTE: EMR is an add-on feature that must be configured to interact with Synergy. Contact LABORIE Service for information on EMR. NXT Owner’s Manual NXT-UM01, MAN2010...
Synergy Software Patient Details The Patient Details section (Figure 30) in the Main Window allows the user to: View previously conducted studies under the Studies tab • View EMR status: report successfully sent or report not sent • Create or edit patient information under the Info tab •...
Synergy Software Database Management Navigate to the Toolbar and select Settings > Database Management. From this option select Archive Patients, Import patients, or Deleted Patients to launch the Database Management window. 4.3.3.1 Archiving Patients Patient data can be archived into an encrypted file on the computer system or on an external hard drive.
Click the Restore button; a User Notification window pops-up, click the OK button to restore the selected patients or the Cancel button to exit. NOTE: To permanently delete patients in the Database management window, contact Laborie Service Department. NXT Owner’s Manual NXT-UM01, MAN2010...
Synergy Software 4.4 Study Management Workflows Synergy Workflows control the structure and content of each study. Each Workflow has standard Workflow Steps: Procedure, Questionnaire, Report, and Review. Workflows vary by the phases contained in the default configuration of their Procedure workflow step. Access Workflow settings in the Workflow Configurator window by navigating to Settings >...
Noise Reduction Level, Video Settings, and Profile Settings. i. Use the Urocap Noise Reduction Level feature to detect and smooth chart spikes caused by environmental noise. Contact LABORIE Service for guidance on option selection. ii. Activate or disable Video for custom workflows and set Desired Frame Rate.
Synergy Software Urogynecology 2P Urodynamics or Urogynecology 3P Urodynamics. Alternatively create a custom workflow by adding a UPP phase to the default workflow templates provided. b. The Phases setting tab allows for customization of auto and manual event markers per phase and of phase layout. i.
Synergy Software 4.4.2.1 Workflow Step Settings Tabs The following settings tabs are available under the Procedure Workflow Step: Global: Graph Settings, Channel Settings, Other Settings, EMG Settings, Video • Settings, Profile Settings Channels: Channel Name, Channel Position, Channel Type, Unit, ScaleMax, •...
Synergy Software Click the Phases tab on the Workflow Configuration Window. Select a phase to customize from those listed. Scroll down to the control panel section located at the bottom of the window. Refer to Figure 40. Figure 40: Workflow Configuration Window - Control Panel Verify that the desired number of Control Panel buttons are provided.
Synergy Software To customize the appearance of buttons, click the Options icon located beside the button. A window will appear where minor button functions and appearance may be adjusted (Figure 42). Options include button caption, colors, and connectivity. Once edits are complete click OK. Figure 42: Edit Control Panel Button Click the OK button located at the top of the Workflow Configuration window to implement all changes to the custom workflow.
Synergy Software Start Study Button The Start Study button launches Synergy into the Preview Study phase. Figure 44: Start Study Button Once all study inputs are received the button will shift to blue. When greyed out the button cannot be activated. Hover the curser over the Start Study button to activate tool tip functionality which will suggest troubleshooting options, as shown in Figure 45.
Synergy Software To proceed through the workflow steps and their phases, utilize the icons described below in Table 6: Icon Button Description Allows the user to proceed to the next phase of the Next study. During a Urodynamic study the ‘Next’ icon will be accompanied by the title of the next study phase.
Synergy Software 4.5.1.2 Consumable Traceability During the Preview phase, the user will need to enter each of the consumables applicable for the study under the Consumable Traceability window. • The user can do this by either scanning the SmartSense tag on the product using the registration systems embedded in the Cart’s docking station and the PIM NXT, or by manually entering the required information.
Synergy Software a. To use the scanner, hold the SmartSense tag located on the consumable’s packaging in proximity to the scanner located on the NXT Docking Station for 3 seconds. Verify that the consumable appears in the Consumable Traceability window. b.
Synergy Software In the Flow and Voiding phases, the Synergy software will automatically start to • record data when the Urocap NXT detects the start of the flow. The Voiding phase will stop 60 seconds after voiding ends by default. The user may also click the Next icon to proceed to the next step.
Synergy Software Select the patient from the Patients Section in the Main Window, refer to Figure 29 for a visual reference. Open the report from the Patient Details section. The report will automatically generate in conformance to the UDS report configuration. Press the copy button located beside the Configuration title (Figure 49).
Synergy Software Enter any comments in the window and select the save icon on the Comments window’s toolbar. Ensure the Comments option is checked under Report content in the Configuration settings. To edit comments from the Configuration Settings, open the Comments dropdown menu and select the click to enter text! option. This will open the Comments window.
Synergy Software 4.5.4.4 Normal Values Normal Values is a data field available for activation within report customization. The Normal Values field is intended to provide an expected range of values for different events which may occur during the study (Figure 52). By default, the Normal Values field will be unchecked in the Configuration Settings for the report.
Synergy Software Click the Select Images icon to launch the Select Images/Burst for Report window. Videos captured will be deconstructed as images in the Image Burst tabs and all single images captured will be found under the Images tab. Refer to Figure 53 below. Figure 53: Select Images/Burst for Report Window Select the images required for the report from each tab.
Synergy Software 4.5.4.6 Sending Reports to EMR Open the report from the Patients section or from the Reports Workflow Step and click the Send to EMR button on the report ribbon bar. This icon will only be visible if the user has an EMR license enabled. When the Send to EMR button is clicked, Synergy validates if the EMR service is running.
Page 66
Synergy Software While in the Review window, the user can: Access, edit, and review content in the Slide Side Bar. • Review UDS Intelligence Score, saved video/images in the Video Panel, and Quick Results. • Edit patient details and patient history. Select images for inclusion in the report.
Synergy Software UDS Intelligence Quality Control (QC) The UDS Intelligence capabilities within Synergy allow the software to provide feedback on user adherence to required standards and procedures throughout the study. The Quality Control window can be accessed by selecting the Quality Control Icon the Slide Sidebar at any time during the study.
Synergy Software 4.6.1.2 Protocol Guidance During the Recording phases of the Procedure workflow step, the Quality Control window will display Protocol Guidance. Two protocol fields are provided: Total Score and Phase Score (Figure 58). The Quality Control icon will shift colors as the phase score evaluation changes.
Synergy Software QC Score and QC Score Summary The QC Score is an aggregated estimate of the quality of the workflow tasks performed while conducting a study. The QC Score can be viewed in a summary at the end of study Report or from the Quality Control icon in the Slide Sidebar.
QUAT wipes (Sani-Cloth or equivalent) • CAUTION: LABORIE has not evaluated the use of cleaning/disinfection reagents which do not appear on the lists above. Use of unapproved reagents could damage the NXT system and should be avoided. For questions on the compatibility of a reagent which is not listed above, contact LABORIE.
Maintenance and Service 5.1.1.2 Cleaning and Disinfection Frequency Cleaning – After each use, all devices should be visually inspected and cleaned • to remove any debris or soiling. Disinfection – Devices should be disinfected in accordance with the frequency • defined below.
Urodynamic study. If dirt or debris is observed in the transducer channels as a part of daily inspection, Contact LABORIE for service and/or disposal instructions. If contamination is observed near the transducer channels during or post procedure, follow...
Maintenance and Service 5.1.3.2 Cleaning the Urocap NXT This section provides instructions for cleaning the Urocap NXT and its accessories. Ensure that the vent plug is inserted into the open vent hole, then follow the instructions provided in the Performing Manual Cleaning section on page to clean the Urocap NXT, being certain not to apply excessive force to the top of the Urocap NXT.
Maintenance and Service Wipe the area with a dry, lint free cloth. Ensure the area is dry to the touch. Reattach, if removed, and close the transducer doors. The transducer doors should only be kept open when reprocessing the area normally covered by the transducer doors.
Perform visual inspection of all units for corrosion, discoloration, pitting, or cracking. Contact LABORIE for service and/or disposal instructions if the unit is observed with these defects. NOTE: There are no rinsing steps required for the cleaning or disinfection processes.
Storage Conditions. For long term storage, LABORIE recommends that the Roam NXT and the Urocap NXT be fully charged prior to storing. Remove the module from the applicable docking station and press the “Reset” button on the module to cause the device to go into long-term sleep mode.
Inspect the Roam NXT and PIM NXT to ensure there are no signs of visible deterioration of the device prior to using them for a Urodynamics Study. General guidelines for caring for these devices are provided below: If a transducer door or belt clip breaks, is damaged or lost, contact LABORIE • Service for a replacement.
Contact LABORIE Service for repair or replacement. 5.2.2.1 Safely Moving the Cart Place the cart into the LABORIE recommended transport position prior to moving it by following the instructions below. Secure loose parts and accessories. Slide the mouse tray into the cart. Close and secure all drawers and doors (as applicable).
Maintenance and Service Maneuver the cart by grasping the cutout area on the keyboard tray with two hands. Always maintain a clear visual of the path forward. Figure 63: Moving the NXT Pro in Transport Position CAUTION: To prevent any risk of damage to the cart: Do not lean on trays, accessories, monitors, the main tower, or the spine of any •...
Maintenance and Service 5.3 Verifying Calibration LABORIE recommends users verify calibration annually, or when any module has been shaken, dropped, or transported between sites, to confirm system accuracy. IMPORTANT: NXT System modules do not require user calibration as the system modules are provided calibrated.
Maintenance and Service d) Press the Zero Weight button. e) The Absolute Weight and Relative Weight for Urocap NXT should measure at 0.0g +/- 2g immediately following zeroing. To verify calibration, check the absolute and relative weight with water. 5.3.1.2 Checking the Absolute and Relative Weight with Water To complete calibration verification for the Urocap NXT, check the absolute and relative weight with water:...
Maintenance and Service 5.3.2.1 Verifying EMG Capability To begin verifying calibration for Roam NXT EMG capability, gather one calibrated signal generator (capable of generating sinusoidal waves at 10Hz-200 µV) with one alligator cable, three Touchproof 1.5mm connectors, and one EMG input cable. To complete calibration verification: a) Setup a signal generator;...
Maintenance and Service 5.3.2.2 Verifying Pressure Measurement Capability To begin verifying calibration for the PIM NXT transducers, gather one Cylinder at least 25 cm tall and one T-DOC Air Charged Catheter. To complete calibration verification: a) Fill a cylinder with 20cm of water. b) Place a T-DOC Air Charged Catheter inside the cylinder so that the tip of the catheter is touching the bottom of the cylinder.
Maintenance and Service 5.3.3.1 Checking the Total Weight without a Saline Bag To begin calibration verification, measure the total weight of the Infusion Pump NXT and Infusion Volume Transducer without a saline bag: a) Setup the Infusion Volume Transducer without a saline bag. b) Click the Settings icon in the title bar, then click System Settings >...
Refer to the diagram below in Figure 70 for a visual representation of the method to isolate the Urocap NXT from supply mains when configured with a wall charger. For further information regarding electrical isolation, contact LABORIE Service. Direct electrical contact...
Maintenance and Service 5.5 Computer Virus Protection Each computer purchased from LABORIE is verified to be virus-free prior to shipment and is installed with an antivirus program. It is the customer’s responsibility to correctly use and maintain the antivirus program. Customers may use an alternative antivirus program at their own risk.
Plug system into a known working the devices? outlet? electrical outlet. Damaged power cord? Unplug system and contact LABORIE for replacement power cord. Power cord not Ensure power cord is secure at the connected properly? base of system and at the electrical outlet.
Page 89
Maintenance and Service Symptoms Possible Cause(s) Corrective Action(s) EMG response is Electrodes not picking up Move either one or both measuring too high and the proper muscle electrodes (the ones attached to the channel saturated? group? red & black leads) closer together. EMG reading is too Mains frequency pickup? Check the measurement electrodes and...
Page 90
Uroflowmeter before starting the study. Funnel is touching the Adjust commode chair or beaker? Uroflowmeter. Incorrect beaker in use? Use beakers supplied by LABORIE only. Bluetooth Connection Unable to connect Connection broken? Reduce the distance between the via Bluetooth? device and the computer. The...
Page 91
For more information on custom workflows refer to the Study Management Workflows section on page 50. If problems continue, contact the LABORIE Service team at 1-800-333-1039 or email service@laborie.com. NXT Owner’s Manual, NXT-UM01, MAN2010...
Maintenance and Service Software Error Messages If the following error messages are displayed in Synergy software, contact the LABORIE Service team at 1-800-333-1039 or email service@laborie.com: Hardware Failure • Device is not ready for another process • Data could not be written to EEPROM •...
Hardware Hardware AIO Computer For information on the specifications of the All-In-One PC (AIO PC) provided with the NXT System, refer to Table 13: Equipment Specifications in the Appendix section beginning on page 106. Bluetooth Dongle NXT The Bluetooth Dongle NXT, as shown in Figure 71, allows modules such as the Roam NXT and Urocap NXT to wirelessly communicate with the NXT system.
Hardware Key Icon Function Key Icon Function Begin Recording Pump Speed Increase Stop Active Pump Speed Hardware Decrease First Sensation Pump On/Off Marker First Desire to Balance/Equalize 1 Void Marker (Standard) Normal Desire to Void Balance/Equalize 2 Marker Maximum Capacity Zero All Pressures Marker Cough Marker...
Hardware Remote Control NXT Battery Replacement To replace the Remote Control NXT battery: Use a Philips screwdriver to unscrew the two screws on the battery door at the back of the Remote NXT. Remove the used battery and replace it with the new battery. Ensure that there is correct battery contact orientation.
Hardware For information on setting up the pump tubing, refer to the Infusion Pump Tubing Setup section on page 22. 6.5 Infusion Volume Transducer The infusion volume transducer is a part of the NXT Infusion Hook. The transducer allows Synergy to measure the fluid used to infuse the bladder during a Urodynamic study. The infusion volume transducer (Figure 75) is setup as follows: Hang a 1000 ml bag of sterile saline or water on the NXT Infusion Hook.
Hardware 6.6 Roam NXT The Roam NXT is a wireless capable module intended to be connected to the detachable PIM NXT (Figure 76). Together, the devices are used to record pressure and EMG during Urodynamic study. Figure 76: PIM NXT Connected to Roam NXT The PIM NXT is the connection point for patient pressure and EMG inputs.
Hardware Replacing the PIM NXT The PIM NXT must be connected to the Roam NXT to facilitate functionality of both devices. Follow the instructions provided below for replacing the PIM NXT. IMPORTANT: The PIM NXT is not intended to be interchangeable during regular use. When the PIM NXT is directly connected to the Roam NXT it should only be detached for cleaning or maintenance.
Roam NXT is completely drained, the device will not beep when inserted into the docking station. Table 11 indicates the different statuses for the Roam NXT and its battery. LABORIE recommends for the user to have the device plugged in, or for the device to have at least 1- 2 hours of battery charge before beginning a test.
Hardware Solid Amber: Low (Battery capacity Amber: Not ready to use. sufficient for at least 1 hour of Return the Roam NXT to docking recording) station for charging. Flashing Amber: Very Low (Battery capacity sufficient for at least 1/2 an hour of recording) Lack of Illumination: Empty/Sleep Lack of Illumination: Device inactive mode (No declared battery...
Table 12: Urocap NXT LED Signals LABORIE recommends for the user to have at least 1-2 hours of battery charge on the Urocap NXT prior to the start of a test or to complete the test with the Urocap NXT plugged in using the NXT Magnetic Pogo Pin Cable.
Hardware Urocap NXT Charging Hub The Urocap NXT Charging Hub serves as a docking station for the Urocap NXT. The Urocap NXT cannot be utilized for Urodynamic study while the device is seated in the Urocap NXT Charging Hub. To complete a Urodynamic study while the Urocap NXT is charging, use the NXT Magnetic Pogo Pin Cable to connect the Urocap NXT to the Urocap NXT Charging Hub.
Hardware 6.8 NXT Pro Cart The NXT Pro Cart (Figure 85), acts as the backbone for the NXT system. Features of the NXT Pro Cart include: • Four-inch swivel castor wheels for curb clearance with brakes for stability. • An adjustable monitor mount (130mm height, +25°/-15° tilt, 180° swivel adjustment, 90°...
Hardware 6.9 T-DOC Air-Charged Catheters The T-DOC NXT Air-Charged Catheter series are part of the next-generation LABORIE systems designed for use during Urodynamic studies (Figure 87). Figure 87: T-DOC NXT Air-Charged Catheter The T-DOC Air-Charged Catheters are single-use catheters intended for •...
EMG Electrodes The following EMG patches are required for the user to perform a Urology Urodynamic study: Snap-on electrodes (ELE428) and cable (Compatible LABORIE Cables) • For information on how to place the electrodes, refer to the Placing the EMG Patch Electrodes section on page 25.
(-10° C to +40° C, 20% to 80% RH Non- Condensing) are labelled on the outside of the packaging. Laborie tests to more extreme conditions to ensure this product can withstand any unexpected conditions which may arise. Laborie has not performed testing to verify system functionality when transported outside of the limits provided NXT Owner’s Manual, NXT-UM01, MAN2010...
Page 107
Appendix Temperature: -20° C to +40° C Storage Conditions Humidity: up to 85 ± 5% RH Non-Condensing IMPORTANT: LABORIE has not performed testing to verify system functionality when stored outside of the limits provided. All-in-one PC Computer: All-in-One medical-grade PC P/N: POC-W243 Screen: 24", 1920 x 1080...
Page 108
Type BF Applied Part: N/A External Materials: Dow Corning QP1-250, Cycoloy C6200-111, Elastosil 3003/30 A/B, Gold-plated Brass, Loctite 401 & 242 IMPORTANT: LABORIE cannot guarantee the IP47 rating unless the vent plug is inserted into the open vent hole. Remote NXT...
Appendix Appendix B: Symbols Glossary Authorized Representative in the European Community Consult Instructions for Date of Manufacture (5.1.2) SGS certification - Use (5.4.3) Read Operator’s Manual Manufacturer (5.1.1) (5.1.3) certified to U.S. and Indicates the (M002) Manufacturer Indicates the device Canadian safety Indicates the date of manufacturer’s device...
Appendix Protective Earth, Not for general Do Not Stack Grounding waste. This product (2402) (Table D.1, 7) is designated for Indicates the items separate collection Identifies any General Warning Sign shall not be vertically at an appropriate (W001) terminal intended for Safety Label, Safe Safety Label, Safe stacked, either...
EMC information provided in this manual. For EM compliance, it is important to only use catheters and accessories as supplied or recommended by LABORIE. Contact LABORIE for approved catheters. Recommendations for actions to assure that the NXT remains safe regarding EM disturbances: The equipment should be visually inspected regularly for damaged cables and connectors.
Page 113
Appendix Guidance and Manufacturer’s Declaration—Electromagnetic Immunity Immunity IEC 60601 Test Compliance Level Electromagnetic Test Level Environment —Guidance Electrostatic ±8 kV contact ±8 kV contact Floors should be wood, discharge concrete or ceramic tile. If ±15 kV air ±15 kV air (ESD) floors are covered with synthetic material, the...
Page 114
Appendix Guidance and Manufacturer’s Declaration—Electromagnetic Immunity Immunity test IEC 60601 Test Compliance Level Electromagnetic level Environment—Guidance Portable and mobile RF communications equipment should be used no closer to any part of the NXT, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Appendix 1970 MHz 28 V/m 2450 MHz 9 V/m 5240/5500/ 5785 MHz Note 1: At 80 MHz and 800 MHz, the higher frequency range applies. Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. a.
Appendix Recommended Separation Distances between Portable and Mobile RF Communications Equipment and the NXT The NXT is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the NXT can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the NXT as recommended below, according to the maximum output power of the communications equipment.