SmartPrep

Specifications

The SmartPrep degasser has been designed and tested to meet the specifications
provided in the following table.
Degas Method
Number of Stations
Temperature Range
Accuracy
Soak Time
Ramp and Soak Cycles
Ramp Rate
Electrical
Voltage
Power
Frequency
Environment
Temperature
Humidity
Gases Used
Flow Rate
Physical
Height
Width
Depth
Weight
Sep 98
Table 1-1. SmartPrep Specifications
Characteristic
Specification
Flowing gas
Six
o
Ambient to 425
C, in 1
station is controlled separately
±5
o
C of set point for temperatures up to 100
±10
o
C for temperaures between 100 and 300
±15
o
C for temperatures over 300
0 to 999 minutes, in 1-minute increments
Up to 5 ramp and soak cycles
o
o
From 5
C/min. to 20
85 to 265 VAC
500 VA Maximum
50/60 Hz
o
10 to 35
C, operating; 0 to 50
20 to 80% relative, noncondensing
Nitrogen and other inert gases
3
0 - 500 cm
/min.
50 cm (20 in.)
50 cm (20 in.)
40 cm (16 in.)
15 kg (33 lbs)
Specifications
o
C increments; each
o
o
o
C
C/min.
o
C, nonoperating
C;
C;
1-3