SmartPrep
500- An error occurred while loading the application control information.
Data entry cannot be performed. (Code #%d)"
Cause:
Action:
502- File cannot be opened for writing. It is already in use.
Cause:
Action:
503 File <name> cannot be analyzed. It is currently being edited.
Cause:
Action:
504- Error accessing the sample information file <name>.
Cause:
Action:
506- The file <name> cannot be prepared for analysis. It is open for editing
and contains errors.
Cause:
Action:
Aug 98
An error occurred accessing the control information disk file re-
quired by this application.
The disk drive may have failed or be corrupt. Run diagnostics
on the disk drive.
The file you are attempting to access is in use by some other
part of the application.
Close the window or application which is currently using the de-
sired file and try again.
You are trying to start a degas operation using a file that is
open for editing.
Finish editing the file, save and close it, then start the degas op-
eration.
An unexplained error prevents you from accessing this file.
The hard disk drive may be corrupt. Run diagnositics.
You are trying to start a degas operation using a file that con-
tains errors and is open for editing.
Finish editing this file, save and close it, then start the degas op-
eration.
Error Messages