Specifications
Specifications
The Model 1305 MultiVolume Pycnometer has been designed and tested to meet
the specificaitons provided below.
150 cm
(indicates 35 to 150 cm
35 cm
(indicates 5 to 35 cm
5 cm
(indicates less than 5 cm
——————— ACCURACY/REPRODUCIBILITY ———————
Depends upon nature of sample. Samples containing water vapor or other vola-
tile matter will adversely affect results. Extremely high specific surface area
samples may require a computed correction. Clean, dry samples of medium to
low surface area will run as follows:
150 cm
35 cm
5 cm
1-6
Characteristic
————————— SAMPLE VOLUME —————————
3
Range:
Cylindrical sample chamber, 5.080 cm (2.000 in.)
diameter x 7.620 cm (3.000 in.) usable depth;
3
)
allows up to 154.4 cm
cup measures 4.572 cm (1.800 in.) inside
diameter x 7.277 cm (2.865 in.) usable depth,
holds up to 119.5 cm
other material.
3
Range:
Sample chamber insert with a cylindrical sample
chamber, 3.556 cm (1.400 in.) diameter x 3.556
3
)
cm (1.400 in.) usable depth; allows up to 35.32
cm
cm (1.341 in.) inside diameter x 3.429 cm (1.350
in.) usable depth holds up to 31.25 cm
volume of powders or other material.
3
Range:
Precision sample chamber insert with a
cylindrical sample chamber, 1.842 cm (0.725 in.)
3
)
diameter x 1.842 cm (0.725 in.) usable depth;
allows up to 4.905 cm
cup measures 1.689 cm (0.665 in.) inside
diameter x 1.755 cm (0.691 in.) usable depth
holds up to 3.933 cm
other material.
3
±0.2% (±0.3 cm
Range:
±0.1% or better usually attained
±0.2% of full scale (±0.070 cm
3
Range:
±0.1% (±0.035cm
±0.2% of full scale (±0.010 cm
3
Range:
±0.1% (±0.005cm
Model 1305 MultiVolume Pycnometer
Specification
3
nominal volume. Sample
3
bulk volume of powder or
3
nominal volume. Sample cup measures 3.406
3
nominal volume. Sample
3
bulk volume of powder or
3
) of full scale guaranteed;
3
) usually attained
3
) usually attained
3
bulk
3
) guaranteed;
3
)guaranteed.
Jun 99
Need help?