Decimal point (DECIMAL P.)

Set the decimal point of torque to display.
<Setting range>

Sign (SIGN)

Set the sign of torque.
<Setting range>
- NORMAL
- REVERSE
- ABSOLUTE

Zero calibration (ZERO CAL.)

Zero point is registered while torque load is not applied to UTMⅡ. Furthermore, if the initial torque
remains in the jig and so on for the purpose of measurement, the zero point may be registered while
the initial torque is being applied; however, the initial torque range needs to be subtracted from the
signal input range.
<Setting range>
1.
Press the
the setting value display screen.
2.
Display the "CONFIG TORQUE" screen using the
3.
Select zero calibration using the
4.
Unload the sensor and press the

Minimum scale (MIN. SCALE) (optional if there is no change)

Set the minimum unit of torque such as scale interval and scale division.
<Setting range>
17

(These items only available in case the UTM capacity is set to OTHER)

0000, 000.0, 00.00, 0.000
NORMAL, REVERSE, ABSOLUTE
Viewing the UTMⅡ from the drive side, the direction in which the load side is
twisted counterclockwise is displayed as the positive direction.
Viewing the UTMⅡ from the drive side, the direction in which the load side is
twisted clockwise is displayed as the positive direction.
Shows the absolute value of torque without the direction of the load.
-5.000 to 5.000 [V]
SET
key on the main screen and enter
keys.
1, 2, 5, 10, 20, 50, 100
keys.
SET
key.