6.0 Paddles Test (If applicable)

Tools Needed: QED 6 Defibrillator Analyzer.
Test Setup: If applicable, connect the universal cable to the paddles and place the paddles in paddle wells.
Do this...
6.1
Turn the selector switch to DEFIB (For AED
units, turn the selector switch to ON and
select Manual mode.) Press and hold the
ENERGY DOWN button on the sternum paddle.
6.2
Press and release the ENERGY UP button on
the sternum paddle for each setting.
6.3
Press and release the RECORDER button on
the sternum paddle.
6.4
Press and release the RECORDER button
again.
6.5
Select 30J using the paddle ENERGY button.
Press the CHARGE button on the Apex paddle.
6.6
Press and release the APEX SHOCK button.
6.7
Press and release the STERNUM SHOCK
button.
6.8
Press and hold both paddles SHOCK buttons.
Record your results on the Maintenance Tests Checklist.
1-16
Verify that...
The energy selection decreases to 1J.
The joules selection increases as follows: Biphasic 1-10, 15, 20, 30, 50, 75,
100, 120, 150, 200J.
Monophasic (DSW): 1-10,15, 20, 30, 50, 75, 100, 150, 200, 300, 360.
The recorder starts printing.
The recorder stops printing.
The unit charges to 30J, then the red LED charge indicator illuminates and the
charge tone sounds.
Note: The front panel shock button does not illuminate.
The unit does not discharge.
The unit does not discharge.
The unit discharges. The TEST OK message displays and the red LED turns
off. If configured, the recorder prints a strip chart.
Pass/Fail/NA
o o
o o
o o o
o o
o o
o o
o o
o o
9650-0450-01 Rev. U