3. Unplug spade connectors, and remove choke assembly.
5. Remove plastic spacer and disconnect 2 spade connectors on
capacitor.
Note:
During reassembly, wires are marked with polarity, make sure they
match capacitor.
5-26
4. Bleed the excess voltage using a resistor with values of
approximately 5k ohms, 25 watts for 10-20 seconds.
Measure the voltage on the capacitor terminals.
6. After discharging capacitor, short the 2 leads together with
a paperclip until reassembly.