ZOLL M Series Service Manual page 14

Hide thumbs Also See for M Series:
Table of Contents

Advertisement

M Series Service Manual
EtCO
cable, and CAPNOSTAT 3 Mainstream cable with airway adapter, or CAPNOSTAT 3 Sidestream cable with cannula (if option is
2
installed).
Novametrix Capnostat Simulator Yb 1265/7100
Paddles (if used).
)
Fluke Biomedical BP Pump 2 NIBP Monitor Analyzer (For NIBP units only) with NIBP cable and cuff (if NIBP option is installed), or
Fluke Biomedical Cufflink Analyzer (if NIBP option is installed)
IBP Temperature Simulator 9100-0402-TF
DNI Nevada 214B Patient SImulator, with M Series interface cable
The Fluke Biomedical BP Pump NIBP Monitor Analyzer and the Fluke Biomedical BP Pump NIBP Analyzer use different technologies for testing
NIBP monitors and therefore, the manual provides two different procedures for performing the NIBP Volume Leak test with each of these types of
test equipment.