Test Setup:
Remove red lead from power supply and connect to 0.1ohm resistor.
Connect other end of resistor to "+" terminal of power supply using a second red label.
Connect multimeter across the resistor.
Set voltage scale (if DVM is not autoranging) to 220mV.
Do this...
4.7
System Current Test
Set power supply to 10.3V.
9650-0450-01 Rev. U
Red
+
Battery Well
_
Black
Verify that...
_
DMM
+
+
15 Amp
Supply
_
M Series Service Manual
Pass/Fail
1-13