ZOLL M Series Service Manual page 33

Hide thumbs Also See for M Series:
Table of Contents

Advertisement

Note:
The universal cable should not be connected to any equipment at the beginning of this test.
Do this...
11.1
Turn the selector switch to DEFIB mode.
(For AED units, turn the selector switch to ON
and select Manual mode.)
Set leads to PADS.
11.2
Connect the universal cable to the MFC test
port.
11.3
Select energy level of 100J and press the
CHARGE button.
11.4
Press the SHOCK button.
11.5
Set energy level to 30J.
11.6
Press the CHARGE button.
11.7
Press and hold SHOCK button.
Record your results on the Maintenance Tests Checklist.
9650-0450-01 Rev. U
Verify that...
The unit displays the CHECK PADS/POOR PAD CONTACT message.
DEFIB PAD SHORT message displays.
The charge time is greater than 2 seconds and less than 10 seconds and the
unit displays the message, SELECT 30J FOR TEST.
The unit does not discharge. The unit displays the message, DEFIB PAD
SHORT.
The unit discharges internally.
The unit charges to 30J and displays message, DEFIB 30J READY. The
charge ready tone sounds.
The unit discharges and displays the number of joules delivered and the TEST
OK message. For example, the message at the top of the printed strip chart
would read as follows:
Biphasic Units: 30 JOULES TEST OK. TEST_CUR=10-14A DEFIB_IMPED=0.
Monophasic Units: 30 JOULES TEST OK, 37-47A, JOULES DEL=30,
IMPED=0
The impedance value may range from 0 to 5 ohms.
If configured, the recorder prints a strip chart.
M Series Service Manual
Pass/Fail
o o
o o
o o
o o
o o
o o
o o
1-23