15.0 Advisory Message Test

Tools Needed: QED 6 Defibrillator Analyzer.
Test Setup:
1. Connect the universal cable via the adapter (D.N.I #3010-0378).
2.
Attach the M Series to the QED 6 Defibrillator Analyzer.
Do this...
15.1
Turn the selector switch to DEFIB mode.
(For AED units, turn the selector switch to ON.)
15.2
Select VF (ventricular fibrillation) on the QED 6,
then press the ANALYZE button.
15.3
Press the SHOCK button.
15.4
Select the NSR (normal sinus rhythm) on the
simulator, then press the ANALYZE button.
Record your results on the Maintenance Tests Checklist.
9650-0450-01 Rev. U
Verify that...
The unit displays the following messages:
ANALYZING ECG.
STAND CLEAR.*
SHOCK ADVISED.*
PRESS SHOCK.*+
*AED's audio prompts are standard. Advisory audio prompts are user
configurable.
+If configured for auto charge.
The unit discharges.
The unit displays the following messages:
ANALYZING ECG.
STAND CLEAR.*
NO SHOCK ADVISED.*
*AED's audio prompts are standard.
M Series Service Manual