Non-Invasive
Blood
Pressure
9650-0214-01 Rev. D

Advertisement

Table of Contents
loading

Summary of Contents for ZOLL M Series

  • Page 1 Non-Invasive Blood Pressure 9650-0214-01 Rev. D...
  • Page 2 The issue date or revision level for this operation guide is shown on the front cover. ZOLL and M Series are trademarks of ZOLL Medical Corporation.  2002 by ZOLL Medical Corporation. All rights reserved.
  • Page 3: Table Of Contents

    ..........................16 AINTAINING THE PTION Performing the Daily Checkout Procedure....................16 Cleaning the Hose............................17 Cleaning Non-Disposable Cuffs ........................17 ................................17 PECIFICATIONS ..............................19 OFTWARE ICENSE ............................. 19 RDERING CCESSORIES ..............................21 ROUBLESHOOTING ZOLL M Series Non-Invasive Blood Pressure...
  • Page 4 (This page intentionally left blank.) ZOLL M Series Non-Invasive Blood Pressure...
  • Page 5: Preface

    NIBP option. Consult the Troubleshooting section if the M Series fails to operate as expected. If you do not know how to use the M Series, see the M Series Operator’s Guide and the relevant inserts. Thoroughly read the safety considerations and warnings sections in both the M Series Operator’s Guide and the relevant inserts before operating your M Series product.
  • Page 6: Cuff And Hose Safety

    Contraindications for Use Use of the ZOLL M Series NIBP option is not indicated for use on neonatal patients or infants whose upper arm circumference is less than 13 cm. SunTech CE Contact...
  • Page 7: Introduction

    • Perform any of the above operations without leaving Monitor, Pacer or Defib mode. How Do I Use the NIBP Option? To take safe and accurate blood pressure measurements using the M Series NIBP option, you must do all of the following: 1.
  • Page 8: How Does The Nibp Option Work

    6. Blood pressure values display on the screen. This cycle takes approximately 30 seconds. If set to do so, the M Series repeats the cycle at a user- selected interval. If the sensor cannot detect systolic pressure, the M Series immediately increases the cuff inflation pressure and completes the measurement.
  • Page 9: How Do I Read The Display

    NIBP Display Area Softkey Labels Softkeys NIBP Connector NIBP Button NOTE Depending on the options included in your M Series, your front panel and display may differ slightly from the illustrations in this insert. ZOLL M Series Non-Invasive Blood Pressure...
  • Page 10 The M Series default unit of measure is millimeters of mercury (mmHg), but you can configure the M Series to display kilopascals (kPa). See the M Series Configuration Guide for instructions on how to configure alternate units of measure.
  • Page 11: Preparing To Take Measurements

    Before taking NIBP measurements, select the appropriately sized cuff, connect the hose to the M Series unit and to the cuff, and apply the cuff to the patient. Additional cuffs and hoses can be ordered from ZOLL Medical Corporation. See ”Ordering Accessories” on page 19 for part numbers.
  • Page 12: Applying The Cuff To The Patient

    Use the default settings unless they are clearly inappropriate for the patient. Any changes to these settings remain in effect until either the settings are changed or 10 seconds after the M Series is turned off. If you have not received training on setting NIBP features, do not use the NIBP option.
  • Page 13: Displaying The Nibp Menu

    1. Press the Param softkey. If the Param softkey label is not displayed, press the Return softkey until “Param” displays. The M Series displays the Parameter menu (see the following figure): Figure 3: Parameter Menu 2. Press the Select softkey until “NIBP” highlights.
  • Page 14: Selecting Cuff Inflation Settings

    140). Use the default settings unless they are clearly inappropriate. Any changes to these settings remain in effect until either the settings are changed or 10 seconds after the M Series is turned off (returning the settings to their default).
  • Page 15: Setting Cuff Inflation Pressure

    30-120 mmHg; 4.0-16.0 kPa When the M Series is turned on, all alarm functions are disabled and the alarm limits are set to their default values. Any changes to these settings remain in effect until either the settings are changed or 10 seconds after the M Series is turned off.
  • Page 16: Setting Alarm Limits

    30 mmHg (4.0 kPa) and completes the measurement. If after 180 seconds the M Series has not been able to determine all three blood pressure values or find a fault, it aborts the measurement cycle and deflates the cuff.
  • Page 17 WARNING If the M Series unit takes a measurement but indicates the presence of artifact in the signal (denoted by “*” in the NIBP Display area), the measurement may not be accurate. Under such circumstances, perform additional blood pressure measurements. If you repeatedly obtain artifact, use alternate techniques to obtain blood pressure prior to taking clinical action.
  • Page 18: Taking A Single Measurement

    Taking a Single Measurement You can take a single measurement at any time by pressing the NIBP button, unless the M Series is charged or charging. CAUTION Make sure that no measurement is being taken when you start a new measurement. If you press the NIBP button while the M Series is taking a measurement, the measurement aborts immediately and the cuff deflates.
  • Page 19: Starting Stat Measurements

    The M Series continues taking automatic measurements until you press the NIBP Auto softkey, or the unit is turned off for more than 10 seconds. If the M Series is turned off for less than 10 seconds, it resumes automatic measurements as scheduled. You can take an additional measurement between automatic measurements, by pressing the NIBP button.
  • Page 20: Starting Automatic Measurements

    2. Press the Dec or the Inc softkey to select measurement interval. 3. Press the Return softkey to return to NIBP menu. The measurement interval sets and remains until 10 seconds after the M Series is turned off or the value is reset.
  • Page 21: Stopping All Automatic Measurements

    STAT measurement, the M Series terminates all measurements. If you press the NIBP button while the M Series is taking an automatic measurement, it aborts the current measurement, deflates the cuff, and continues all subsequent measurements at the set interval.
  • Page 22: Printing Data

    Printing Data See the “Recorder Operation” section of the M Series Operator’s Guide for instructions on how to record data. If you have already taken NIBP measurements, press the RECORDER button to print a strip chart that includes the following values for the currently displayed measurement: •...
  • Page 23: Cleaning The Hose

    American National Standard, Electronic or automated sphyg- momanometers (AAMI SP10). To receive a copy of the report containing the AAMI SP10 test results, contact ZOLL Technical Support at (800) 348-9011 or (781) 229-0020...
  • Page 24 High 6.7 – 26.7 kPa; Low 25 – 100 mmHg Low 3.3 – 13.3 kPa Mean High 60 – 220 mmHg; High 8.0 – 29.3 kPa; Low 30 – 120 mmHg Low 4.0 – 16.0 kPa ZOLL M Series Non-Invasive Blood Pressure...
  • Page 25: Software License

    Grant of License: In consideration of payment of the software license fee which is part of the price paid for this product ZOLL Medical Corporation grants the Purchaser a non-exclusive license, without right to sublicense, to use the system software in object- code form only.
  • Page 26 Table 13: NIBP Cuff Connector Accessories Connectors: Part Number: Cuff hose connector (female) 0310-0026 ZOLL M Series Non-Invasive Blood Pressure...
  • Page 27: Troubleshooting

    The following table lists the NIBP messages, the problem each message indicates, and the actions you should take to fix the problem. Read this section carefully before monitoring patients. For further assistance, call ZOLL Technical Service at 1-800-348-9011. International customers should call the nearest authorized ZOLL Medical Corporation distributor.
  • Page 28 Reapply cuff or select smaller cuff. Patient’s limb below level of heart. Raise patient’s limb to heart level. False low reading Cuff too large. Select smaller cuff Limb above level of heart. Lower patient’s limb to heart level. ZOLL M Series Non-Invasive Blood Pressure...
  • Page 29 ECG-out jack vi alarm 9 cuff placement vi auto adjust 8 cuff selection vi cuff inflation adjustment 8 M Series general v cuff inflation pressure 8, 9 cleaning caution vi hose kinking vi measurement interval 13 Maintaining NIBP 16...
  • Page 30 range 17 caution 17 reading on display 4 temperature 18 Operating conditions Suspended alarm vi Measurement temperature 18 30 second safety delay 10 SYSTEM FAULT message Operating pressure accuracy 17 significance of 22 Pressure automatic 13 Systolic Blood Pressure operating 18 cycle time 18 alarm limit 18 delay after 10...