ZOLL M Series Service Manual page 75

Hide thumbs Also See for M Series:
Table of Contents

Advertisement

Error Message
RELEASE BUTTONS
RELEASE SHOCK
REPLACE BATTERY
REPLACE EtCO
2
SENSOR
REPORT FULL
REPORT HALTED
RESERVED 1
RETRY ANALYSIS
SELECT 30J FOR TEST
SELECT DEFIB MODE
SELECT LIMB LEADS
SELECT PADS
9650-0450-01 Rev. U
Explanation
Simultaneous external paddle button presses detected before
unit reached full defib charge (ready state).
Discharge switch(es) closed when pressing charge button.
Discharge button pressed before defib reached ready state.
Battery voltage is less than absolute minimum. Shutdown
imminent.
EtC0
SENSOR WARM UP message displays for more
2
than five minutes. Sensor defective.
Summary report memory full.
Summary report stops printing unexpectedly.
The watchdog timer is not functional in the unit.
Advisory message in conjunction with noisy ECG. Analysis
halted.
Attempt to run a self test at an energy other than 30J.
Analyze button pressed in pace or monitor mode.
Paddles or augmented ECG leads selected when continuous
analysis active or started.
Lead I, II, or III selected when analyze pressed.
User
Technical Action
Advisory
Release buttons.
Release shock button.
Check paddles.
Replace controls board.
Replace with charged battery.
Replace sensor cable.
Erase summary report.
Turn unit off and then back on
again. Print Summary again. If
fault persists, replace system
board.
Turn off unit and then turn on to
reset. If fault persists, replace
system board.
Select limb leads I, II, III or
MFE
M Series Service Manual
2-15