ZOLL M Series Service Manual page 36

Hide thumbs Also See for M Series:
Table of Contents

Advertisement

M Series Service Manual
Do this...
13.10
Press the SHOCK button.
13.11
Press the ENERGY SELECT up arrow until the
unit displays 200J.
13.12
Press the CHARGE button. Wait for the
SHOCK button to illuminate.
13.13
Press the SHOCK button.
13.14
(Monophasic Units) Press the ENERGY
SELECT up arrow unit the unit displays 300J.
13.15
Press the CHARGE button. Wait for the
SHOCK button to illuminate.
13.16
Press the SHOCK button.
13.17
(Monophasic Units) Press the ENERGY
SELECT up arrow unit the unit displays 360J.
13.18
Press the CHARGE button. Wait for the
SHOCK button to illuminate.
13.19
Press the SHOCK button.
13.20
(Biphasic Units) Press the CHARGE button and
start timing with a stopwatch. Stop timing when
the SHOCK button lights.
13.21
Press the SHOCK button.
13.22
(Monophasic Units) Press the CHARGE button
and start timing with a stopwatch. Stop timing
when the SHOCK button illuminates.
1-26
Verify that...
The unit discharges 83J-122J (monophasic 85J-115J) into the simulator.
The unit displays the message, DEFIB 200J SEL.
The unit displays the message, DEFIB 200J RDY.
The unit discharges 182J-247J (monophasic 170J-230J) into the simulator.
The unit displays the message, DEFIB 300J SEL.
The unit displays the message, DEFIB 300J RDY.
The unit discharges 255J-345J into the simulator.
The unit displays the message, DEFIB 360J SEL.
The unit displays the message, DEFIB 360J RDY.
The unit discharges 306J-414J into the simulator.
The charge time is between 3-6 seconds.
On the strip chart, the Patient Current is between 20-24 Amps, and the Defib
Impedance is between 46-54 Ohms.
Note: This will occur if configured to Auto-Generate Strips on event.
The charge time is between 4-8 seconds.
Pass/Fail/NA
o o o
o o o
o o o
o o o
o o o
o o o
o o o
o o o
o o o
o o o
o o o
o o o
o o o
9650-0450-01 Rev. U