5. If present remove EtCO
2
connector panel by lifting straight out.
Note:
During reassembly of the SPO
clockwise and route cable between the molex connections.
7. Lift HV module assembly straight up and out of the device.
9650-0450-01 Rev. U
and SPO
connectors from the
2
connector, turn connector once
2
6. Cut tie wrap holding HV cable to connector panel. Then
remove HV cables from wire loop.
Note:
During reassembly, make sure tie wrap is positioned toward the
bottom of connector panel as shown.
8. Remove HV isolator, and unplug MFC signal cable.