Synchronized Cardioversion Test - ZOLL M Series Service Manual

Hide thumbs Also See for M Series:
Table of Contents

Advertisement

M Series Service Manual

12.0 Synchronized Cardioversion Test

Tools Needed: QED 6 Defibrillator Analyzer.
Test Setup:
1. Connect the universal cable via the adapter (D.N.I #3010-0378) to the defibrillator analyzer.
2.
Set ECG on analyzer to 60 -120 BPM.
Do this...
12.1
Press LEAD button to select PADS and Size
X1.
12.2
Press the SYNC softkey on the defibrillator.
Enter synchronized cardioversion timing test
mode on the defibrillator analyzer.
12.3
Select 200J (360J for Monophasic units), and
then press the CHARGE button. When the
SHOCK button lights, press and hold the
SHOCK button.
Record your results on the Maintenance Tests Checklist.
1-24
Verify that...
Sync markers appear on display.
Note: Sync markers display on the monitor. The sync marker appears as a
down arrow over the ECG R-wave peaks on strip chart and display.
On the analyzer display, the R-wave to shock delay is less than60
milliseconds.
Pass/Fail