1.0 Physical Inspection of the Unit

Tools Needed: Battery.
Test Setup: None.
Observe this...
1.1
Housing
Is the unit clean and undamaged?
1.2
Does the unit show signs of excessive wear?
Does the handle work properly?
1.3
1.4
Does the recorder door open and close properly?
Are input connectors clean and undamaged?
1.5
1.6
Are there any cracks in the housing?
1.7
Do the front panel or selector switches have any damage or cracks?
Are there any loose housing parts?
1.8
1.9
Cables
Are all cables free of cracks, cuts, exposed or broken wires?
1.10
Are all cable bend/strain reliefs undamaged and free of excessive cable
wear?
1.11
Paddles
Do the adult and pedi plates have major scratches or show signs of damage?
1.12
Do the adult shoes slide on and off easily to expose the covered pedi plates?
Are the paddles clean (e.g., free of gel) and undamaged? (if applicable)
1.13
1.14
Battery
Place battery in battery well.
9650-0450-01 Rev. U
Pass/Fail/NA
o o
o o
o o
o o
o o
o o
o o
o o
o o
o o
o o o
o o o
o o o
o o