5.0 Leakage Current Test

Tools Needed: See the manufacturer's instructions or supplied specifications for the leakage tester you use.
Test Setup: See the manufacturer's instructions or supplied specifications for the leakage tester you use. Repeat leakage test with accessories: MFC,
external paddles, and anterior/posterior paddles.
Maximum Leakage Acceptance Limits
Normal Condition
ECG
10
MFC
100
Earth
500
*Single fault considered AC mains on applied part.
9650-0450-01 Rev. U
Single Fault Condition*
50
100
1000