Installing The Source Block Assembly - Thermo Scientific Surveyor MSQ Plus Hardware Manual

Table of Contents

Advertisement

Installing the Source Block Assembly

Thermo Scientific
Figure 72. View of the source block with the proper alignment of the extraction cone
insulator, extraction cone, and hexapole insulator
You must reinstall the source block assembly before you prepare the system for operation.
CAUTION The corona pin is very sharp. Do not attempt to install the entrance cone or
the cone wash nozzle before you turn the corona pin knob to its vertical position. See
Figure
73.
To install the source block assembly into the MS detector
1. Insert the source block assembly (P/N FM102573) into the source block enclosure, and
then tighten the thumbscrews on the source block. See
CAUTION The tip of the cone wash nozzle is very fragile.
2. Handling the cone wash nozzle by its base because the tip is fragile, insert the cone wash
nozzle (P/N FM102521) into the source block, and then carefully rotate the nozzle tip to
the right. See
Figure
73.
5
Routine and Preventive Maintenance
Maintaining the Source Block Assembly
Properly aligned components
Figure 67
on
page
85.
Surveyor MSQ Plus Hardware Manual
91