Daily Operation

Things to Do Before Operating the MS Detector

Checking the Nitrogen Supply

Thermo Scientific
To optimize the performance of your Surveyor MSQ Plus MS detector, you need to perform
various routine operations both before and after you operate the system.
Contents
Things to Do Before Operating the MS Detector
Things to Do After Operating the MS Detector
This section contains the following procedures that you might want to perform on a daily
basis before you begin your analyses:
Checking the Nitrogen Supply
Checking the Disk Space
Checking the Oil Level in the Oil Mist Filter
Check the nitrogen supply on the regulator of the nitrogen gas tank or liquid nitrogen boil-off
tank. Make sure that you have sufficient gas for your analysis. Typical nitrogen consumption
is 15800 liters per day (based on a 24 hr day). If necessary, replace the tank. Verify that the
pressure of nitrogen reaching the mass spectrometer is between 520 ± 35 kPa (75 ± 5 psi).
If necessary, adjust the pressure with the tank pressure regulator.