Routine and Preventive Maintenance
Maintaining the APCI Probe

Removing the APCI Capillary

68
Surveyor MSQ Plus Hardware Manual
5. Turn the locking plate clockwise to the open position and remove the APCI probe from
the probe heater. See
Figure
Figure 48. Surveyor MSQ Plus MS detector set up for the APCI mode
Figure 49. Remove the APCI probe from the probe heater
To remove the capillary from the APCI probe
1. Using a 2.5 mm Allen key, unscrew the M3 × 10 cap head stainless steel screw
(P/N FM103046) from the probe clamp. See
2. Remove the probe clamp and the locking plate.
49.
Figure
50.
PEEK fingertight fitting
Locking plate
(in locked position)
Probe heater
Locking plate
(in open position)
Probe heater
Thermo Scientific