Turning the MS Detector Off from the Tune Window

Thermo Scientific
Figure 77. Info view - Status page, showing the shortcut menu for the Surveyor MSQ Plus
MS detector
To turn off the Surveyor MSQ Plus MS detector from the Tune window
1. If you are using the optional cone wash pump, turn it off by turning its power switch to
the Off position.
2. Turn off the flow from your LC pump:
a. If your LC pump is controlled from the Xcalibur data system, turn it off from its
respective Direct Control dialog box or place it in the Standby mode from the Info
View - Status page.
b. If your LC pump is not controlled from the Xcalibur data system, turn it off from its
keypad control panel.
3. If the Tune window is not open, open it:
a. Double-click the Tune icon on the Windows desktop.
After you double-click the Tune icon, the Server icon appears in the Windows
taskbar.
b. Double-click the Server icon to open the Tune window.
4. If the Per Method Parameters table is not open, open it by clicking the Per Method
Parameters slide bar on the right side of the Tune window.
6
System Shutdown, Startup, and Reset
Placing the System in the Off Mode
Turning off
the MS detector
Surveyor MSQ Plus Hardware Manual
99