Routine and Preventive Maintenance
Maintaining the APCI Probe

Installing the APCI Probe

72
Surveyor MSQ Plus Hardware Manual
Figure 54. Enlarged view of the tip of the APCI probe
To install the APCI probe into the probe heater
CAUTION Take care when inserting the APCI probe into the probe heater. The capillary
protruding from the end of the probe is fragile and easily damaged. In addition, you can
contaminate the capillary if you let it touch the inner surface of the probe heater as you
insert the probe into the probe heater.
1. Turn the locking plate of the probe clockwise to the open position, and then carefully
slide the APCI probe into the probe heater. See
2. Turn the locking plate counterclockwise to the closed position to lock the probe in place.
See
Figure 48
on
page
3. Connect the PEEK fingertight fitting to the APCI probe. See
10 mm
68.
0.41 mm
depth of APCI spacer
0.5 mm
Figure 49
on
page
68.
Figure
48.