Functional Description
Data System

Tune Window

40
Surveyor MSQ Plus Hardware Manual
Right-clicking the Server icon opens the shortcut menu shown in
allows you to do the following:
• Choose Manual Tune to open the Tune window, where you can optimize the
performance of your Surveyor MSQ Plus MS detector for a specific application.
• Choose Instrument Tune and Calibration to display the Instrument Tuning and
Calibration wizard.
• Choose Vent to turn off the turbomolecular pump and vent the vacuum. If the system is
not under vacuum, the menu contains the Pump command. Choose Pump to pump
down the system. The Edwards forepump is not turned off by venting the system. To turn
off the Edwards forepump without turning off the MS detector, set the On/Off switch on
the forepump to the Off position.
• Choose Exit to close the server. If Xcalibur is running, exiting the server is not allowed
and the error message shown in
Figure 28. Server shortcut menu
Figure 29. Error message displayed when Xcalibur is running and you attempt to exit the Server
application
The Surveyor MSQ Plus Tune window allows you to optimize the MS detector parameters
(manually tune) for your analytes and acquire data to a raw file (.raw). When you finish
tuning your MS detector for a compound of interest, you save the current values of the tuning
parameters in a tune file (.tune). In ESI, you empirically determine the optimal probe
temperature, needle voltage, and cone voltage for each application. In APCI, you empirically
determine the optimal probe temperature, corona current, and cone voltage for each
application.
Figure 29
appears.
Figure
28. This menu