Installing The Source Housing Drain Insert And Sweep Cone - Thermo Scientific ISQ EC Operating Manual

Hide thumbs Also See for ISQ EC:
Table of Contents

Advertisement

5
Installation
5.5.5

Installing the Source Housing Drain Insert and Sweep cone

Page 98
Your instrument ships with the source housing drain insert and sweep cone
uninstalled due to risk of shipping damage to sensitive parts. Please follow
this step-by-step guide prior to turning on the power.
WARNING—Instrument Damage
Wear gloves and eye protection to avoid contamination of sensitive
parts.
AVERTISSEMENT—Risque d'endommagement de l'instrument
Portez des gants et des lunettes de protection pour éviter de contaminer
les parties sensibles.
1)
Open the front bezel by grasping the door on the left edge.
2)
Lift up the top hatch of the instrument to provide sufficient space to
remove the source housing as shown in Figure 43.
Figure 43: Removal of the source housing
3)
Remove PEEK capillary fitting connection at grounding union as
shown in Figure 44.
ISQ EC and ISQ EM Mass Spectrometer Operating Manual

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Isq em

Table of Contents