Restarting The System Following A Complete Shutdown - Thermo Scientific Surveyor MSQ Plus Hardware Manual

Table of Contents

Advertisement

Restarting the System Following a Complete Shutdown

Thermo Scientific
Figure 78. View of taskbar showing the Server icon and the shortcut menu
7. Exit Xcalibur. Then, close the Server.
8. Wait for approximately 2 minutes. Then, turn off the Surveyor MSQ Plus MS detector
by setting the MAINS IN switch to the Off position.
Turning off the power to the Surveyor MSQ Plus MS detector also turns off the power to
the Edwards forepump, which gets its line power from the Pump Out receptacle on the
rear panel of the MS detector.
9. Close the nitrogen gas cylinder at the main regulator.
Following a long-term shutdown, carry out the visual checks listed in the Pre-switch on
checklist shown in
Table
This section contains the following topics:
Checking the System Connections
Restarting the Surveyor MSQ Plus System
After you complete these procedures, the system is ready for a Full System Autotune. Refer to
the Surveyor MSQ Getting Started manual for details.
6, and then follow the system start-up procedure.
6
System Shutdown, Startup, and Reset
Restarting the System Following a Complete Shutdown
Surveyor MSQ Plus Hardware Manual
101

Advertisement

Table of Contents
loading

Table of Contents