3.4.7

Adjusting the Ph Module Position

It is necessary to center the annular diaphragm of
the Ph module.
Center the annular diaphragm using the centering
telescope. Remove one eyepiece, attach the
centering telescope, and then focus on the ring
diaphragm image by turning the eyepiece section of
the centering telescope.
Turn the centering screws of the module using a
specified tool to overlay the annular diaphragm
image on the phase ring image in the objective.
Annular
diaphragm
centering screws
Pull out the
slider to the
front.
HNA condenser slider
Annular diaphragm centering screw
Annular
diaphragm
Centering of the annular diaphragm
47
Chapter 3 Usage of Components
Condenser turret
Annular diaphragm
centering screw
(both sides)
Phase ring