Chapter 7 Assembly of the Devices
7.18

Moving the Microscope

When moving the microscope, attach the TI-BCH carrying handles.
The TI-BCH carrying handles are support handles that are used to move the microscope a short distance,
such as when changing the room layout or when placing the microscope into a cart.
Notes on moving the microscope
Before moving the microscope, thoroughly read
"Safety Precautions" at the beginning of this manual, and always heed all warnings and cautions written therein.
1.
Remove all devices and cables that can be
removed from the microscope.
2.
Remove the caps from the two M6 upper
service threads out of four on each side of
the microscope main body (base).
Service threads are located on the right and left
side of the main body. Remove the two caps
from each side.
3.
Mount the four carrying handles into the
service threads on the microscope main
body (base).
Two service threads are located on the right
and left side of the main body. Mount carrying
handles on both service threads on each side.
Be sure to screw the TI-BCH carrying handles
all the way in.
CAUTION "10. Cautions on carrying the microscope" in
Mounting the carrying handles
176
Service threads
Carrying handle