2.2
Diascopic Phase Contrast (Ph) Microscopy (Continued)
3
Perform Ph microscopy.
3-1 Fully open the dia aperture diaphragm.
[g] Aperture diaphragm open/close lever (
3-2 Adjust the size of the dia field diaphragm
image.
[h] Field diaphragm dial (
3-3 When using an objective with a correction
collar, adjust the correction collar.
[b] Objective (
3.10.1)
Adjust the correction collar of the objective according
to the thickness of the prepared specimen slide or the
cover glass of the culture vessel.
3-4 Move the stage and turn the focus knobs to
observe the sample.
[i]
Stage handle (
3.8.3)
[c] Focus knob (
3.11.1)
3-5 When using another objective, repeat the
procedure from step
3.4.5)
3.3.3)
2
.
19
Chapter 2 Microscopy