Maintenance
Test point for
measuring
sensors
130
The operating hours counter must only be reset when the filter
has been changed.
Reset
Yes
-
The operating hours counter will be reset for the new filter.
No
-
The counter will not be reset.
Select an option using the and arrow buttons.
Confirm your selection with OK.
Performance check
Adequate processing performance must be regularly confirmed by
the user.
The sensor test point for validation is located at the front right on the
top of the machine under the lid or the countertop. To reach the
access point, the lid of the machine must be removed or the machine
must be pulled out from under the countertop.
Open the door.
Unscrew the retaining screws.
Then remove the safety screws on the back of the machine from
the lid and lift the lid to remove it.
Or
Pull the machine out by approx. 6" (15 cm) from under the
countertop.
Yes
No