Problem
Conductivity level too high:
Actual value: µS/cm Max
value: µS/cm
Conductivity level exceeds
limit
Conductivity module requires
calibration
Conductivity module
communication error
Spray pressure exceeds
tolerance
Spray pressure fluctuating too
much
Possible cause and solution
Carry-over of conductive substances during reprocessing.
Check the process.
Empty or faulty water softener or demineralization system.
Check external water softener or demineralization
systems.
Reactivate the systems if necessary.
Work on the on-site water supply.
Contact a suitably qualified plumber.
Plumbing connections transposed.
Observe the markings on the plumbing connections (see
"Connection to the water supply").
If the level drops below the measuring range, the
conductivity cannot be determined.
Contact Miele Service.
The conductivity meter must be recalibrated.
Contact Miele Service.
The connection to the conductivity meter has been
interrupted.
Contact Miele Service.
The wash pressure differs from the reference value.
Possible causes of fluctuations in the wash pressure
include:
- defective water connections,
- open adapters,
- foam build-up.
Identify and resolve the cause of this.
The program is not interrupted. Nevertheless, the wash
load must be reprocessed.
A program was interrupted because of severe fluctuations
in the wash pressure.
Possible causes of fluctuations in the wash pressure
include:
- defective water connections,
- open adapters,
- foam build-up.
Identify and resolve the cause of this.
Reprocess the load again.
Problem solving guide