Adding liquid detergents and neutralizers
Message "Check dispensing system
[X]"
The running program is interrupted.
^ Check the container(s) and
dispensing hoses shown in the
display.
[x]: Instead of an X, the number of the affected
dispensing system will appear.
^ Refill or replace the container(s)
shown in the display.
^ Use the appropriate service program
to prime the affected dispensing
system (clear it of air).
The level query for dispensing
systems which are not being used
can be turned off to avoid fault
messages in the display (see the
Programming Manual, "Machine
function / Dosage/Container lance").
52
Priming the dispensing system
Whenever a container has been
allowed to completely empty, it must be
primed (cleared of air) after refilling.
^ Select the corresponding service
program, e. g.
^ Press the h start button.
,
The cancelled program must be
repeated.
Fill DOS1
.